Background: Scanning electron microscopy (SEM) studies found that the porosity of thrombi might vary among individuals. However, its relationship with the clinical presentation and efficacy of mechanical thrombectomy (MT) remains unknown. We aimed to characterize the ultrastructure of thrombi and explore its association with the complexity of MT and clot perviousness.
Methods: SEM was used to observe the morphological features of different components of thrombi obtained from patients with anterior circulation large vessel occlusion undergoing MT and to determine the porosity of thrombi by semi-quantitative analysis. Non-porous thrombi were defined as thrombi with porosity <2%. Clot perviousness was also evaluated using thrombus attenuation increase on CT perfusion (TAI). We assessed the complexity of MT by attempts of retrieval >3 and procedural duration >60 min, defined as the time interval between groin puncture and recanalization.
Results: A total of 49 thrombi were analyzed and 31 (63.3%) were classified as non-porous thrombi. The presence of non-porous thrombi was negatively associated with procedure >60 min (OR 0.152, 95% CI 0.031 to 0.734, p=0.019) and attempts >3 (OR 0.194, 95% CI 0.046 to 0.822, p=0.026) after adjustment. Additionally, receiver operating characteristic curve analysis indicated that TAI <17.9 Hounsfield units could predict the presence of non-porous thrombi with an area under the curve of 0.915.
Conclusions: Non-porous thrombi on SEM are easier to be retrieved during MT and could be identified as less pervious clots on CT images.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1136/neurintsurg-2022-018689 | DOI Listing |
Int J Biol Macromol
December 2023
Department of Inorganic Polymers, "Petru Poni" Institute of Macromolecular Chemistry, Aleea Gr. Ghica Voda 41 A, Iasi 700487, Romania. Electronic address:
Multicomponent composites based on natural biopolymers: chitosan, starch and gelatin in two different ratios (0.5:1:1 and 1:1:1) were in situ crosslinked by intermolecular interactions and used as matrices for zinc oxide and magnetite fillers. The bionanocomposite films have been evaluated by spectral and microscopy methods: Fourier-Transform Infrared spectrometry (FT-IR), Scanning Electron Microscopy (SEM) and Atomic Force Microscopy (AFM) confirming the electrostatic and hydrogen bonding interactions between the components of the polymeric matrices and the inorganic fillers and the crosslinking process.
View Article and Find Full Text PDFJ Neurointerv Surg
February 2023
Department of Neurology, the Second Affiliated Hospital of Zhejiang University, School of Medicine, Hangzhou, China
Background: Scanning electron microscopy (SEM) studies found that the porosity of thrombi might vary among individuals. However, its relationship with the clinical presentation and efficacy of mechanical thrombectomy (MT) remains unknown. We aimed to characterize the ultrastructure of thrombi and explore its association with the complexity of MT and clot perviousness.
View Article and Find Full Text PDFJ Control Release
August 2020
Utah Center for Nanomedicine, Nano Institute of Utah, and University of Utah, Salt Lake City, Utah, United States; Department of Pharmaceutics and Pharmaceutical Chemistry, Nano Institute of Utah, and University of Utah, Salt Lake City, Utah, United States; Department of Biomedical Engineering, University of Utah, Salt Lake City, Utah, United States. Electronic address:
Chronic toxicity evaluations of nanotechnology-based drugs are essential to support initiation of clinical trials. Ideally such evaluations should address the dosing strategy in human applications and provide sufficient information for long-term usage. Herein, we investigated one-year toxicity of non-surface modified silica nanoparticles (SNPs) with variations in size and porosity (Stöber SNPs 46 ± 4.
View Article and Find Full Text PDFJ Cardiovasc Surg (Torino)
August 2003
Department of Surgery, University of California, San Francisco, CA 94143, USA.
The aim of this study was to explain variations in the results of endovascular aneurysm repair as a consequences of device design. Low profile, trackable systems, such as Zenith and Excluder, rarely fail to traverse the iliac arteries, even in the presence of iliac tortuosity or stenosis. In most patients, optimal sizing is only possible with systems, such as Zenith, Talent, and Quantum lp, that have a wide range of diameters.
View Article and Find Full Text PDFInterv Neuroradiol
November 2000
Department of Neurosurgery, Kitano Medical Institute and Hospital; Osaka, Japan.
The use of stents improves the result after balloon coronary angioplasty. Restenosis due to neointimal hyperplasia and proliferation of smooth muscle cells are, however, a concern. In the present report, we studied the prevention of restenosis to allow endothelial cell migration and growth to proceed through micropores using our developed stent graft with micropored segmented polyurethane (SPU) thin film in a normal beagle model.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!