Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Patient-specific instrumentation (PSI) may potentially improve humeral osteotomy in shoulder arthroplasty. The purpose of this study was to compare the deviation between planned and postosteotomy humeral inclination, retrotorsion, and height in shoulder arthroplasty, using PSI vs. standard cutting guides (SCG).
Methods: Twenty fresh-frozen cadaveric specimens were allocated to undergo humeral osteotomy using either PSI or SCG, such that the 2 groups have similar age, gender, and side. Preosteotomy computed tomography (CT) scan was performed and used for the 3-dimensional (3D) planning. The osteotomy procedure was performed using a PSI designed for each specimen or an SCG depending on the group. A postosteotomy CT scan was performed. The preosteotomy and postosteotomy 3D CT scan reconstructions were superimposed to calculate the deviation between planned and postosteotomy inclination, retrotorsion, and height. Outliers were defined as cases with 1 or more of the following deviations: >5° inclination, >10° retrotorsion, and >3 mm height. The deviation and outliers in inclination, retrotorsion, and height were compared between the 2 groups.
Results: The deviations between planned and postosteotomy parameters were similar among the PSI and SCG groups for inclination (P = .260), whereas they were significantly greater in the SCG group for retrotorsion (P < .001) and height (P = .003). There were 8 outliers in the SCG group, compared with only 1 outlier in the PSI group (P = .005). Most outliers in the SCG group were due to deviation >10° in retrotorsion.
Conclusion: After 3D planning, PSI had less deviation between planned and postosteotomy humeral retrotorsion and height, relative to SCG.