This study aimed to evaluate in vitro performance refractive segmented EDOF intraocular lenses under polychromatic light using an optical bench that complies with the ISO 11979-2 Norm. The through focus modulation transfer function (TF-MTF) of the Femtis Comfort LS-313 MF15 (Oculentis GmbH, Berlin, Germany) IOL was evaluated for IOLs with three different base powers. The effect of the asymmetry of the segmented designs was evaluated with 3 different wavelengths and with polychromatic light at a 3.0 mm and 5.0 mm pupil diameter. It was demonstrated that the TF-MTF curves exhibit a bifocal profile that, in practice, results in an EDOF design. As a consequence of the LCA, the TF-MTF values in white light were lower than in monochromatic light. Images of the USAF test chart were obtained to confirm the prediction of the TF-MTFs. We found that Femtis Comfort is a bifocal low-addition IOL and this fact can result in an EDOF effect which was obtained previously in clinical trials. Moreover, we showed that the base power influences the IOL optical quality, which results as more effective for high powers (hyperopic eyes) than for low powers (myopic eyes). The LCA of the segmented refractive design was very low and presumably not clinically relevant.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8954764 | PMC |
http://dx.doi.org/10.3390/jcm11061480 | DOI Listing |
J Vasc Bras
January 2025
Universidade Federal da Paraíba - UFPB, Hospital Universitário Lauro Wanderley - HULW, João Pessoa, PB, Brasil.
Pulmonary arteriovenous malformations (PAVM) are characterized by abnormal pulmonary vessels forming arteriovenous shunts that compromise oxygenation of the blood, causing hypoxemia, and predispose to infections and cerebral ischemia. The patient in this case was a 38-year-old male who presented with tachypnea and dyspnea, cyanosis of extremities, and significant digital clubbing. The patient had structural epilepsy secondary to neurosurgery for a cerebral abscess during childhood.
View Article and Find Full Text PDFOphthalmic Physiol Opt
January 2025
Robert O Curle Ophthalmology Suite, Institute for Regeneration and Repair, The University of Edinburgh, Edinburgh, UK.
Purpose: To determine whether imaging features derived from fundus photographs contain 3D eye shape information beyond that available from spherical equivalent refraction (SER).
Methods: We analysed 99 eyes of 68 normal adults in the UK Biobank. An ellipsoid was fitted to the entire volume of each posterior eye (vitreous chamber without the lens)-segmented from magnetic resonance imaging of the brain.
J Clin Med
January 2025
Ophthalmology Section, Department of Medicine and Surgery, University of Perugia, 06132 Perugia, Italy.
To report the cosmetic, clinical, and visual outcomes of a combined surgical approach for treating a corneal/limbal dermoid using excision and a three-layered amniotic membrane graft with fibrin glue. An 18-year-old female presented with impaired vision and ocular discomfort caused by a prominent dome-shaped limbal congenital dermoid on the inferotemporal cornea, resulting in a significant aesthetic concern. A full assessment, including refraction, best-corrected visual acuity (BCVA), corneal topography, aberrometry and anterior segment OCT (AS-OCT) was conducted to plan the surgical approach.
View Article and Find Full Text PDFJ Cataract Refract Surg
January 2025
Yokohama Sky Building Eye Clinic, 2-19-12 Takashima, Yokohama, Nishi-ku, Kanagawa, 220-0011, Japan.
Purpose: To compare the refractive accuracy of the Barrett True axial length (BTAL) formula, newly integrated into ARGOS, with that of the Barrett Universal II (BUII) formula calculated using axial length (AL) from IOL Master 700.
Setting: Private clinics in Kanagawa, Japan.
Design: Retrospective observational study.
J Cataract Refract Surg
January 2025
Kepler University Hospital, University Clinic for Ophthalmology and Optometry.
Purpose: To comprehensively evaluate the tilt of the crystalline lens in phakic eyes and its relationship with optical biometry parameters.
Setting: The retrospective analysis was conducted at the Department for Ophthalmology and Optometry, Kepler University Clinic in Linz, Austria.
Design: Lens tilt assessment was performed through a combined analysis of optical biometry (IOL Master 700) and anterior segment optical coherence tomography (AS-OCT) data to understand its relationship.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!