  • A novel insulated-tip ablation catheter (SMT) was developed to enhance radiofrequency ablation (RFA) efficiency by directing more energy into targeted tissue, rather than dissipating it into surrounding areas.
  • A comparison study between SMT and conventional RFA catheters (RFC) involved 274 applications using SMT and 74 using RFC, targeting porcine myocardium and thigh muscle preparations.
  • Results indicated that SMT could direct about 66.8% of energy into targeted tissue, producing lesions of similar size at lower power, and larger lesions at higher power, while also resulting in fewer complications like charring and steam pops.

Article Abstract

Introduction: During radiofrequency ablation (RFA) using conventional RFA catheters (RFC), ~90% of the energy dissipates into the bloodstream/surrounding tissue. We hypothesized that a novel insulated-tip ablation catheter (SMT) capable of blocking the radiofrequency path may focus most of the energy into the targeted tissue while utilizing reduced power and irrigation.

Methods: This study evaluated the outcomes of RFA using SMT versus an RFC in silico, ex vivo, and in vivo. Radiofrequency applications were delivered over porcine myocardium (ex vivo) and porcine thigh muscle preparations superfused with heparinized blood (in vivo). Altogether, 274 radiofrequency applications were delivered using SMT (4-15 W, 2 or 20 ml/min) and 74 applications using RFC (30 W, 30 ml/min).

Results: RFA using SMT proved capable of directing 66.8% of the radiofrequency energy into the targeted tissue. Accordingly, low power-low irrigation RFA using SMT (8-12 W, 2 ml/min) yielded lesion sizes comparable with RFC, whereas high power-high irrigation (15 W, 20 ml/min) RFA with SMT yielded lesions larger than RFC (p < .05). Although SMT was associated with greater impedance drops ex vivo and in vivo, ablation using RFC was associated with increased charring/steam pop/tissue cavitation (p < .05). Lastly, lesions created with SMT were more homogeneous than RFC (p < .001).

Conclusion: Low power-low irrigation (8-12 W, 2 ml/min) RFA using the novel SMT ablation catheter can create more uniform, but comparable-sized lesions as RFC with reduced charring/steam pop/tissue cavitation. High power-high irrigation (15 W, 20 ml/min) RFA with SMT yields lesions larger than RFC.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9325397PMC
http://dx.doi.org/10.1111/jce.15461DOI Listing

Publication Analysis

Top Keywords

rfa smt
16
radiofrequency ablation
8
novel insulated-tip
8
insulated-tip ablation
8
ablation catheter
8
energy targeted
8
targeted tissue
8
radiofrequency applications
8
applications delivered
8
radiofrequency
6

Similar Publications

Radiofrequency Ablation of Benign and Malignant Thyroid Nodules.

Radiographics

October 2022

From the Departments of Radiology (P.J.N., S.M.T., A.N.K., R.A.L., M.R.C., B.T.W., J.J.S.) and Endocrinology (M.R.C., M.N.S.), Mayo Clinic, 200 1st St SW, Rochester, MN 55905-0002.

Article Synopsis
  • Thyroid nodule treatment now focuses on personalizing care based on the nodule's cause and the patient's health, with increasing interest in radiofrequency ablation (RFA) as a less invasive alternative to surgery.
  • RFA involves heating the thyroid nodule to cause tissue destruction and is recognized by various medical guidelines as an effective option for benign nodules and certain cases where surgery is not feasible.
  • While RFA is effective and safe for many patients, its use for treating small thyroid cancers remains debated, highlighting the importance of careful patient selection and procedural technique to minimize risks.
View Article and Find Full Text PDF
Article Synopsis
  • A novel insulated-tip ablation catheter (SMT) was developed to enhance radiofrequency ablation (RFA) efficiency by directing more energy into targeted tissue, rather than dissipating it into surrounding areas.
  • A comparison study between SMT and conventional RFA catheters (RFC) involved 274 applications using SMT and 74 using RFC, targeting porcine myocardium and thigh muscle preparations.
  • Results indicated that SMT could direct about 66.8% of energy into targeted tissue, producing lesions of similar size at lower power, and larger lesions at higher power, while also resulting in fewer complications like charring and steam pops.
View Article and Find Full Text PDF

Purpose: To identify the predictive value of demographic and clinical factors for determining changes in physical and depressive symptom among hepatocellular carcinoma (HCC) patients over time.

Methods: We performed a prospective cohort study of 128 patients newly diagnosed with HCC in Taiwan. Each patient had four time-point data after the follow-up.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!