Background: Patellar resurfacing is routinely performed during total knee arthroplasty to reduce pain associated with patellofemoral osteoarthritis. With 3-dimensional ingrowth materials readily available, the present study aimed to evaluate if cemented polyethylene (CP) patellar buttons conferred higher ultimate load to failure than press-fit metal-backed (PF) buttons in axial compression.

Material And Methods: Ten matched cadaveric and 20 composite patellae were resurfaced and implanted with either a PF or CP button. Biomechanical testing using an MTS machine was performed to measure the force required to generate a periprosthetic patella fracture. Mean load to failure and load to failure per 1-mm patellar thickness were compared with a paired and independent samples Students' t-test for the cadaveric and composite patellae, respectively.

Results: The average load to failure for the matched cadaveric patellae with PF implants was significantly lower than that for patellae with CP buttons (4082.05 N vs 5898.37 N,  = .045). The average load to failure for composite patella with PF implants was significantly higher than that for composite patellae with CP implants (6004.09 N vs 4551.40 N,  = .001). The mean load to failure per 1-mm patellar thickness was also significantly higher for composite patellae with PF implants (263.80 N/mm vs 200.37 N/mm,  = .001).

Conclusion: Cadaveric patellae with cemented implants had a significantly higher ultimate load to failure in axial compression than press-fit patella. However, this result was reversed in the composite model. Exploration of biological and composite model properties could provide further insight into patellar implant selection during total knee arthroplasty.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8927789PMC
http://dx.doi.org/10.1016/j.artd.2022.02.012DOI Listing

Publication Analysis

Top Keywords

load failure
28
composite patellae
16
patellae implants
12
total knee
8
knee arthroplasty
8
higher ultimate
8
ultimate load
8
matched cadaveric
8
cadaveric composite
8
failure 1-mm
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!