Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Purpose: We compared the setup errors determined by an optical imaging system (OSIS) in women who received breast-conserving surgery (BCS) followed by whole-breast radiotherapy (WBRT) with those from cone-beam computed tomography (CBCT) carried out routinely.
Methods: We compared 130 setup errors in 10 patients undergoing WBRT following BCS by analyzing the translational and rotational couch shifts via CBCT and OSIS. Patients were treated with intensity-modulated radiotherapy (IMRT). The patient outline extracted from the planning reference Computed tomography (CT) was used as the reference for OSIS and CBCT alignment during treatment. We detected the setup uncertainty using CBCT and OSIS at the first five fractionations of RT and then twice a week.
Results: The absolute translational setup error (mean ± Standard deviation (SD)) in x (lateral), y (longitudinal), and z (vertical) axes detected by the OSIS was 0.14 ± 0.18, 0.15 ± 0.14, and 0.13 ± 0.13 cm, respectively. The rotational setup error (mean ± SD) in Rx (pitch), Ry (roll), and Rz (yaw) axes was 0.77 ± 0.54, 0.76 ± 0.61, and 1.23 ± 0.95, respectively. Significant difference is observed only in one direction (Rx, p = 0.03) in the paired setup errors obtaining from OSIS and CBCT, without significant differences in five directions.
Conclusion: OSIS is a repeatable and reliable system that can be used to detect misalignments with accuracy, which is capable of supplementing CBCT for WBRT after BCS. We believe that an OSIS may be easier to use, quicker, and reduce overall dose as this method of patient alignment does not require ionizing radiation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9121044 | PMC |
http://dx.doi.org/10.1002/acm2.13578 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!