Ambulatory blood pressure (BP) monitoring (ABPM) is vital for screening cardiovascular activity. The American College of Cardiology/American Heart Association guideline for the prevention, detection, evaluation, and management of BP in adults recommends measuring BP outside the office setting using daytime ABPM. The recommendation to use night-day BP measurements to confirm hypertension is consistent with the recommendation of several other guidelines. In recent studies, ABPM was used to measure BP at regular intervals, and it reduces the effect of the environment on BP. Out-of-office measurements are highly recommended by almost all hypertension organizations. However, traditional ABPM devices based on the oscillometric technique usually interrupt sleep. For all-day ABPM purposes, a photoplethysmography (PPG)-based wrist-type device has been developed as a convenient tool. This optical, noninvasive device estimates BP using morphological characteristics from PPG waveforms. As measurement can be affected by multiple variables, calibration is necessary to ensure that the calculated BP values are accurate. However, few studies focused on adaptive calibration. A novel adaptive calibration model, which is data-driven and embedded in a wearable device, was proposed. The features from a 15 s PPG waveform and personal information were input for estimation of BP values and our data-driven calibration model. The model had a feedback calibration process using the exponential Gaussian process regression method to calibrate BP values and avoid inter- and intra-subject variability, ensuring accuracy in long-term ABPM. The estimation error of BP (ΔBP = actual BP-estimated BP) of systolic BP was -0.1776 ± 4.7361 mmHg; ≤15 mmHg, 99.225%, and of diastolic BP was -0.3846 ± 6.3688 mmHg; ≤15 mmHg, 98.191%. The success rate was improved, and the results corresponded to the Association for the Advancement of Medical Instrumentation standard and British Hypertension Society Grading criteria for medical regulation. Using machine learning with a feedback calibration model could be used to assess ABPM for clinical purposes.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8914760PMC
http://dx.doi.org/10.3390/s22051873DOI Listing

Publication Analysis

Top Keywords

feedback calibration
12
calibration model
12
model feedback
8
blood pressure
8
adaptive calibration
8
mmhg ≤15
8
≤15 mmhg
8
calibration
7
abpm
7
data-driven model
4

Similar Publications

A compact dual-gas sensor based on the two near-infrared distributed feedback diode lasers and a multipass cell has been established for the simultaneous measurement of methane (CH) and acetylene (CH). The time division multiplexing calibration-free direct absorption spectroscopy is used to eliminate the cross interference in the application of multicomponent gas sensors. A wavelength stabilization technique based on the proportion integration differentiation feedback control is developed to suppress laser wavelength drift and an H-infinity (H) filter algorithm to reduce the system noise.

View Article and Find Full Text PDF

The development and calibration of a measurement system designed for assessing the performance of the avalanche photodiodes (APDs) used in the Compton scattering polarimeter of the CUSP project is discussed in this work. The designed system is able to characterize the APD gain GAPD and energy resolution across a wide range of temperatures (from -20 °C to +60 °C) and bias voltages Vbias (from 260 V to 410 V). The primary goal was to experimentally determine the GAPD dependence on the and Vbias in order to establish a strategy for stabilizing GAPD by compensating for fluctuations, acting on Vbias.

View Article and Find Full Text PDF

This article reports a 110.2 MHz ultra-low-power phase-locked loop (PLL) for MEMS timing/frequency reference oscillator applications. It utilizes a 6.

View Article and Find Full Text PDF

Background: Preterm birth (PTB) is a leading cause of neonatal mortality, particularly in sub-Saharan Africa where 40% of global neonatal deaths occur. We identified and combined demographic, clinical, and psychosocial correlates of PTB among Kenyan women to develop a risk score.

Methods: We used data from a prospective study enrolling HIV-negative women from 20 antenatal clinics in Western Kenya (NCT03070600).

View Article and Find Full Text PDF

Purpose: This study explored the clinical characteristics of patients with tinnitus who responded to sound therapy and established a predictive model to evaluate the effectiveness of this therapy according to the clinical characteristics.

Methods: A retrospective analysis was performed on 991 subjective tinnitus patients who received compound sound therapy in the Department of Otolaryngology of the local hospital from November 2019 to January 2022.

Results: We found that tinnitus patients with different therapeutic effects had significant differences in the tinnitus side ( = 0.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!