Objective: Sutureless aortic valves are an effective option for aortic valve replacement (AVR) showing non-inferiority to standard stented aortic valves for major cardiovascular and cerebral events at 1-year. We report the 1-year hemodynamic performance of the sutureless prostheses compared with standard aortic valves, assessed by a dedicated echocardiographic core lab.

Methods: Perceval Sutureless Implant vs. Standard Aortic Valve Replacement (PERSIST-AVR) is a prospective, randomized, adaptive, open-label trial. Patients undergoing AVR, as an isolated or combined procedure, were randomized to receive a sutureless [sutureless aortic valve replacement (Su-AVR)] ( = 407) or a stented sutured [surgical AVR (SAVR)] ( = 412) bioprostheses. Site-reported echocardiographic examinations were collected at 1 year. In addition, a subgroup of the trial population (Su-AVR = 71, SAVR = 82) had a complete echocardiographic examination independently assessed by a Core Lab (MedStar Health Research Institute, Washington D.C., USA) for the evaluation of the hemodynamic performance.

Results: The site-reported hemodynamic data of stented valves and sutureless valves are stable and comparable during follow-up, showing stable reduction of mean and peak pressure gradients through one-year follow-up (mean: 12.1 ± 6.2 vs. 11.5 ± 4.6 mmHg; peak: 21.3 ± 11.4 vs. 22.0 ± 8.9 mmHg). These results at 1-year are confirmed in the subgroup by the core-lab assessed echocardiogram with an average mean and peak gradient of 12.8 ± 5.7 and 21.5 ± 9.1 mmHg for Su-AVR, and 13.4 ± 7.7 and 23.0 ± 13.0 mmHg for SAVR. The valve effective orifice area was 1.3 ± 0.4 and 1.4 ± 0.4 cm at 1-year for Su-AVR and SAVR. These improvements are observed across all valve sizes. At 1-year evaluation, 91.3% ( = 42) of patients in Su-AVR and 82.3% in SAVR ( = 51) groups were free from paravalvular leak (PVL). The rate of mild PVL was 4.3% ( = 2) in Su-AVR and 12.9% ( = 8) in the SAVR group. A similar trend is observed for central leak occurrence in both core-lab assessed echo groups.

Conclusion: At 1-year of follow-up of a PERSIST-AVR patient sub-group, the study showed comparable hemodynamic performance in the sutureless and the stented-valve groups, confirmed by independent echo core lab. Perceval sutureless prosthesis provides optimal sealing at the annulus with equivalent PVL and central regurgitation extent rates compared to sutured valves. Sutureless valves are therefore a reliable and essential technology within the modern therapeutic possibilities to treat aortic valve disease.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8894864PMC
http://dx.doi.org/10.3389/fcvm.2022.844876DOI Listing

Publication Analysis

Top Keywords

aortic valve
20
valve replacement
16
hemodynamic performance
12
performance sutureless
12
aortic valves
12
sutureless
9
aortic
8
standard aortic
8
perceval sutureless
8
su-avr savr
8

Similar Publications

Background: Despite advancements in valve implantation devices, vascular access complications (VAC) remain a significant cause of morbidity and mortality for those undergoing transcatheter aortic valve replacement (TAVR). We describe pre-operative imaging analysis of the aortoiliac and femoral arterial beds using the TransAtlantic intersociety consensus (TASC) score, ilio-femoral tortuosity, and procedural characteristics to identify anatomic risk factors predictive of VAC in TAVR.

Methods: Consecutive patients undergoing TAVR from 2012 to 2022 at a single North Dakota hospital were retrospectively reviewed.

View Article and Find Full Text PDF

Background: Transcatheter aortic valve replacement (TAVR) is a well-established treatment option for patients with severe aortic valve stenosis; however, clinical valve thrombosis is a major challenge.

Case Summary: A 92-year-old woman underwent TAVR for severe aortic stenosis. One month later, the patient developed acute heart failure.

View Article and Find Full Text PDF

We report a 75-year-old female with a history of two heart operations: aortic valve replacement (St. Jude Medical 21 mm) at the age of 44 years for severe rheumatic aortic stenosis and mitral valve replacement (Carbomedics 29 mm) at the age of 51 years for rheumatic mitral regurgitation. Decades later, she presented with exertional dyspnea.

View Article and Find Full Text PDF

Introduction The study aimed to retrospectively evaluate the early patient outcome and left ventricular function after mitral valve replacement with a tilting disc valve and total preservation. Patients and methods This retrospective observational study includes patients who underwent mitral valve replacement using a tilting disc valve with total preservation of mitral valvular and subvalvular apparatus from July 2021 to August 2022 at a single center. Results The data were reviewed retrospectively for age, sex, comorbidities, operating time, aortic cross-clamp time, cardiopulmonary bypass time, preoperative and postoperative left ventricular ejection fraction, mean gradient across the mitral valve, left ventricular diameter, left atrial size, atrial fibrillation, complications, mortality, and early patient outcome.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!