Introduction: Grip strength is a reliable biomarker of overall health and physiological well-being. It is widely used in clinical practice as an outcome measure. This paper demonstrates the measurement characteristics of GripAble, a wireless mobile handgrip device that measures grip force both isometrically and elastically-resisted for assessment and training of hand function.
Methods: A series of bench tests were performed to evaluate GripAble's grip force measurement accuracy and sensitivity. Measurement robustness was evaluated through repeated drop tests interwoven with error verification test phases.
Results: GripAble's absolute measurement error at the central position was under 0.81 and 1.67 kg (95 percentiles; N = 47) when measuring elastically and isometrically, respectively, providing similar or better accuracy than the industry-standard Jamar device. Sensitivity was measured as 0.062 ± 0.015 kg (mean ± std; 95 percentiles: [0.036, 0.089] kg; N = 47), independent of the applied force. There was no significant performance degradation following impact from 30 drops from a height >1.5 m.
Conclusion: GripAble is an accurate and reliable grip strength dynamometer. It is highly sensitive and robust, which in combination with other novel features (e.g. portability, telerehabilitation and digital data tracking) enable broad applicability in a range of clinical caseloads and environments.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8891833 | PMC |
http://dx.doi.org/10.1177/20556683221078455 | DOI Listing |
PLoS One
January 2025
Researching in Sport Science: Research Group (CTS-563) of the Andalusian Research Plan, University of Malaga, Malaga, Spain.
The purpose of this longitudinal study was to determine the effects of a rowing training program on physical fitness and body composition in female breast cancer survivors (56.78 ± 6.38 years).
View Article and Find Full Text PDFAnn Med
December 2025
Health Research and Innovatioin Science Centre, Klaipeda University, Klaipeda, Lithuania.
Background: Handgrip strength (HGS) serves as a robust predictor of overall strength across various populations, including individuals with Down Syndrome (DS).
Objective: To analyze the HGS measurement protocols used in studies involving individuals with DS.
Methods: Primary sources were sourced from six databases: PubMed, Scopus, Ovid, Embase, ERIC, and Web of Science, spanning from inception to 23rd December 2023.
Front Psychol
January 2025
School of Nursing, Guangzhou Medical University, Guangzhou, China.
Aim: This study examined the mediating role of resilience and the moderating role of grip strength (handgrip) in the relationship between mindfulness and non-restorative sleep (NRS) and evaluated the non-linear association between mindfulness and NRS among university freshmen students.
Design: This is a cross-sectional descriptive study.
Method: A total of 384 university students were recruited from Be Resilient to Nursing Career (BRNC) from two universities in June 2022.
Quant Imaging Med Surg
January 2025
Department of Ultrasonic Diagnosis, The First Hospital of China Medical University, Shenyang, China.
Background: Patients with essential hypertension have a high risk of muscle mass and strength decline. Ultrasound is a promising method for assessing sarcopenia. This study aimed to analyze the correlation between ultrasound and shear-wave elastography (SWE) features, and muscle mass, muscle strength, and physical performance, and to assess the clinical applicability of ultrasound in the diagnosis of sarcopenia in patients with essential hypertension.
View Article and Find Full Text PDFPhotochem Photobiol
January 2025
Departamento de Fisioterapia, Universidade Federal dos Vales Do Jequitinhonha e, Diamantina, Minas Gerais, Brazil.
This study investigates the effects of photobiomodulation (PBM) with a 660-nm laser on nerve regeneration and muscle morphometry following median nerve axonotmesis in rats. Sixteen Wistar rats were divided into a control group and laser-treated group, with the latter receiving 10 applications of PBM (660 nm; 20 mW; 10 J/cm; 0.4 J; and 20 s) over 2 weeks.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!