Background: Ventricular tachyarrhythmias (VTA) with low and varying signal amplitudes and morphologies may not be successfully identified utilizing traditional implantable cardioverter-defibrillator algorithms.
Objective: Develop and validate a novel algorithm (VF Therapy Assurance, VFTA) to improve detection and timely delivery of high-voltage therapy (HVT) for these arrhythmias.
Methods: Arrhythmia detection was simulated on recorded VTA electrograms (EGMs) utilizing Abbott's Merlin.net database. EGMs where an HVT occurred only when VFTA was enabled, or where VFTA provided an HVT >30 seconds earlier than without VFTA, were readjudicated with physician review. As VFTA never prevents detection or therapy, EGMs where VFTA did not activate or alter HVT were not adjudicated.
Results: Among 564,353 recorded VTA EGMs from 20,000 devices, VFTA altered HVT in 105 EGMs from 67 devices. Physician adjudication determined that 81.9% (86/105) of these EGMs were true undertreated VTA episodes and would have received appropriate HVT with VFTA enabled. Furthermore, 65% of the episodes (56/86) were ventricular fibrillation, were polymorphic, did not self-terminate during the recording window, or were not amenable antitachycardia pacing. Of those, 87.5% (49/56) would not have elicited HVT without VFTA. Overall, VFTA provided new or earlier appropriate HVT in 0.27% (53/20,000) of devices with an increase in inappropriate HVT in 0.07% (14/20,000) devices.
Conclusion: The VFTA algorithm successfully identifies VTA missed by traditional detection algorithms, owing to undersensed ventricular signals resulting in the rate falling below the programmed detection rate. The use of VFTA increases the likelihood of delivering life-saving HVT.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859789 | PMC |
http://dx.doi.org/10.1016/j.hroo.2021.11.009 | DOI Listing |
J Arrhythm
February 2025
Current guidelines recommend cardioverter-defibrillator (ICD) programming, including faster detection rates, longer detection durations, and strict discrimination for supraventricular tachycardia (SVT) to prevent unnecessary ICD treatment. This delayed-style ICD programming could lead to a rise in the possibility of VF undersensing. To avoid this risk, an innovative algorithm known as VF Therapy Assurance (VFTA; Abbott, Sylmar, CA) has been developed.
View Article and Find Full Text PDFNutrients
October 2023
Cardiometabolic Nutrition Group, IMDEA Food, CEI UAM + CSIC, 28049 Madrid, Spain.
Anemia causes hypo-oxygenation in the brain, which could lead to cognitive disorders. We examined dietary iron intake as well as anemia markers (i.e.
View Article and Find Full Text PDFPacing Clin Electrophysiol
February 2023
Department of Cardiology, RTIICS, Kolkata, India.
Newer devices incorporate advanced algorithms to have more accurate delivery of implantable cardioverter-defibrillator (ICD) therapy. Newer models of Abbott Implantable cardioverter-defibrillators like Gallant and Entrant have a new ventricular fibrillation (VF) therapy assurance algorithm. This algorithm has a proven role in avoiding under-detection of VF where some of the smaller electrograms are under-sensed.
View Article and Find Full Text PDFHeart Rhythm O2
February 2022
Abbott, Sylmar, California.
Background: Ventricular tachyarrhythmias (VTA) with low and varying signal amplitudes and morphologies may not be successfully identified utilizing traditional implantable cardioverter-defibrillator algorithms.
Objective: Develop and validate a novel algorithm (VF Therapy Assurance, VFTA) to improve detection and timely delivery of high-voltage therapy (HVT) for these arrhythmias.
Methods: Arrhythmia detection was simulated on recorded VTA electrograms (EGMs) utilizing Abbott's Merlin.
Eur Heart J Cardiovasc Imaging
May 2012
Cardiac Department, National University Heart Center, 1E, Kent Ridge Road, NUHS Tower Block, Level 9, Singapore 119228, Singapore.
Background: In clinical heart failure (HF), inefficient propagation of blood through the left ventricle (LV) may result from suboptimal vortex formation (VF) ability of the LV during early diastole. We aim to (i) validate echocardiographic-derived vortex formation time (adapted) (VFTa) in control subjects and (ii) examine its utility in both systolic and diastolic HF.
Methods: Transthoracic echocardiography was performed in 32 normal subjects and in 130 patients who were hospitalized with HF [91, reduced ejection fraction (rEF) and 39, preserved ejection fraction (pEF)].
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!