Cases of Azygous Coil Extraction.

Heart Rhythm O2

Cardiac Electrophysiology Section, Division of Cardiology, Duke University Medical Center and the Duke Clinical Research Institute, Durham, North Carolina.

Published: February 2022

AI Article Synopsis

  • Two patients with a history of significant heart issues underwent extraction of devices with azygous coils in place, highlighting the lack of existing data on this procedure.
  • Both patients experienced successful removal of the coils using laser technology, with no intraoperative complications reported.
  • The results suggest that azygous coils can be safely removed using standard techniques similar to those used for other lead extractions, despite their varied dwell times.

Article Abstract

Background: Procedural and clinical outcomes of patients undergoing extraction or removal of azygous coils are not well characterized.

Objective: Evaluate outcomes in patients who undergo device extraction with an azygous coil in situ.

Methods: Patients undergoing extraction with an azygous coil in situ between May 2015 and January 2021 were included in this retrospective single-center analysis. Outcomes included procedural success, use of laser and mechanical cutting tools during the procedure, procedural complications, and mortality.

Results: We identified 2 patients undergoing device extraction with an azygous coil in situ with a dwell time greater than 12 months. The patients were male, aged 73 and 83 years. Both had a history of hypertension, atrial fibrillation, heart failure (ejection fractions <15% and 20%), and cardiomyopathy (nonischemic and ischemic), and presented with an infection (case 1 with a single-chamber ICD and Staphylococcus aureus bacteremia, case 2 with a cardiac resynchronization therapy defibrillator pocket infection). The mean dwell time of all 6 leads extracted was 6.43 years (range 1.33-12.63 years), and the 2 azygous coils had dwell times of 1.33 and 6.04 years. In case 1, the azygous coil was inferior to the cardiac silhouette, while in case 2 it was superior. A 14F laser sheath was employed to remove both azygous coils. Both extractions were a complete procedural success in which all leads were removed completely without intraoperative complications.

Conclusion: These cases demonstrate the variable courses of azygous coils, provide proof of concept that they can be removed safely, and illustrate that azygous coils can be removed with the same techniques that are commonly used to remove other types of leads.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8859796PMC
http://dx.doi.org/10.1016/j.hroo.2021.12.004DOI Listing

Publication Analysis

Top Keywords

azygous coil
16
patients undergoing
12
extraction azygous
12
outcomes patients
8
undergoing extraction
8
device extraction
8
coil situ
8
extraction
5
patients
5
cases azygous
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!