Background And Purpose: Linac-based stereotactic radiosurgery (SRS) planning for multi-metastatic cases is a complex and intensive process. A manual planning strategy starts with a template-based set of beam angles and applies modifications though a trial and error process. Beam angle optimization uses patient specific geometric heuristics to determine beam angles that provide optimal target coverage and avoid treating through Organs-at-Risk (OARs). This study expands on a collision prediction application developed using an application programming interface, integrating beam angle optimization and collision prediction into a Stereotactic Optimized Automated Radiotherapy (SOAR) planning algorithm.

Materials And Methods: Twenty-five patient plans, previously treated with SRS for multi-metastatic intracranial tumors, were selected for a retrospective plan study comparing the manual planning strategy to SOAR. The SOAR algorithm was used to select isocenters, table, collimator, and gantry angles, and target groupings for the optimized plans. Dose-volume metrics for relevant OARs and PTVs were compared using double-sided Wilcoxon signed rank tests (α = 0.05). A subset of five patients were included in an efficiency study comparing manual planning times to SOAR automated times.

Results: OAR dose metrics compared between planning strategies showed no statistical difference for the dataset of twenty-five plans. Differences in maximum PTV dose and the conformity index were improved for SOAR planning and statistically significant. The median SOAR planning time was 9.8 min compared to 55 min for the manual planning strategy.

Conclusions: SOAR planning was comparable in plan quality to a manual planning strategy with the possibility for greatly improving planning efficiency through automation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8885579PMC
http://dx.doi.org/10.1016/j.phro.2022.02.009DOI Listing

Publication Analysis

Top Keywords

manual planning
20
soar planning
16
planning
13
beam angle
12
angle optimization
12
planning strategy
12
stereotactic radiosurgery
8
beam angles
8
collision prediction
8
study comparing
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!