Liquid chromatography tandem mass spectrometry (LC-MS/MS) is used routinely in clinical diagnostics; however, automating the sample pretreatment is challenging. We established and evaluated an automated method based on the magnetic bead extraction principle (MBE) to measure normetanephrine (NMN), metanephrine (MN), and 3-methoxytyramine (3-MT). The target analytes were extracted, purified, and concentrated using different solvents and chemical bond-modified magnetic beads transferred via a magnetic bar. The linearity, recovery, matrix effect, and precision of MBE were evaluated thoroughly, and compared with traditional solid-phase extraction (SPE) using 131 plasma samples. The chromatography peaks of metanephrines and 3-MT, extracted via MBE, are symmetrical, without interfering peaks. The linearity was excellent with correlation coefficient (r) > 0.99. The MBE exhibited good reproducibility with within-run coefficient variations (CVs) of 1.96-2.00%, 4.06-5.75%, and 3.89-4.90% for MN, NMN, and 3-MT, respectively. The total CVs for MN, NMN, and 3-MT were 1.96-2.80%, 5.12-5.75%, and 5.44-6.27%, respectively. The relative recoveries for MN, NMN, and 3-MT varied between 93.5 and 107.4%, whereas their biases were all within 10%. The results for MN, NMN, and 3-MT extracted via MBE compared with SPE exhibited excellent correlation, with r > 0.99; the mean bias% for MN, NMN, and 3-MT were small (-2.9%, -3.2%, and -3.2%, respectively). In conclusion, the automated MBE method for measuring plasma metanephrines and 3-MT can be applied in future routine clinical diagnostics, and the MBE principle may indicate a new era for LC-MS/MS in clinical application.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00216-022-03984-x | DOI Listing |
Pract Lab Med
March 2024
Department of Laboratory Medicine, Ren Ji Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200127, China.
Objective: To develop and validate a rapid liquid chromatography-tandem mass spectrometry (LC-MS/MS) method to detect urinary free metanephrines and methoxytyramine, establishing reference intervals.
Methods: Urine samples were diluted with isotope internal standard solution, then analyzed directly using tandem mass spectrometry with multiple reaction monitoring measurement and electrospray ionization source in positive ion mode. Analytical parameters including linearity, lower limit of quantitation, imprecision and accuracy of the method were evaluated.
Molecules
October 2022
Department of Toxicology and Sanitary Chemistry, School of Public Health, Tianjin Medical University, Tianjin 300070, China.
Two different pretreatment approaches have been used for the enrichment and separation of biogenic monoamines and metabolites in plasma for high performance liquid chromatography (HPLC) determination. The first approach, based on on-line packed-fiber solid-phase extraction (PFSPE) coupled with HPLC, allows for the simultaneous detection of epinephrine (E), norepinephrine (NE), dopamine (DA), 3-methoxyl epinephrine (MN), norepinephrine (NMN), 3-methoxytyramine (3-MT), and 5-hydroxytryptamin (5-HT). Using this developed on-line PFSPE-HPLC method, the limit of detections (LODs) of the seven analytes ranged from 1 ng/mL (NMN and MN) to 2 ng/mL (NE, E, DA, 3-MT and 5-HT).
View Article and Find Full Text PDFMolecules
May 2022
Key Laboratory of Laboratory Medical Diagnostics, Ministry of Education, School of Laboratory Medicine, Chongqing Medical University, Chongqing 400016, China.
A novel dual-template magnetic molecularly imprinted polymer (MMIP) was synthesized to extract normetanephrine (NMN), metanephrine (MN) and 3-methoxytyramine (3-MT) from spot urine samples. As the adsorbent of dispersive solid-phase extraction (d-SPE), the MMIP was prepared using dopamine and MN as dual templates, methacrylic acid as the functional monomer, ethylene glycol dimethacrylate as the crosslinking reagent and magnetic nanoparticles as the magnetic core. NMN, MN, 3-MT and creatinine (Cr) in spot urine samples were selectively enriched by d-SPE and detected by HPLC-fluorescence detection/ultraviolet detection.
View Article and Find Full Text PDFAnal Bioanal Chem
May 2022
Department of Clinical Laboratory, Dongcheng District, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, No. 1 Shuaifu Yuan, Beijing, 100730, People's Republic of China.
Liquid chromatography tandem mass spectrometry (LC-MS/MS) is used routinely in clinical diagnostics; however, automating the sample pretreatment is challenging. We established and evaluated an automated method based on the magnetic bead extraction principle (MBE) to measure normetanephrine (NMN), metanephrine (MN), and 3-methoxytyramine (3-MT). The target analytes were extracted, purified, and concentrated using different solvents and chemical bond-modified magnetic beads transferred via a magnetic bar.
View Article and Find Full Text PDFScand J Clin Lab Invest
November 2021
Center of Laboratory Medicine, Key Laboratory for Molecular Diagnostics of Cardiovascular Diseases, State Key Laboratory of Cardiovascular Disease, National Center for Cardiovascular Diseases & Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, Beijing, China.
Background: Accurate diagnosis of pheochromocytoma and paraganglioma (PPGLs) is highly dependent on the detection of metanephrines and catecholamines. However, the systematic investigation on influencing factors including specimen (plasma or whole blood), anticoagulant, storage conditions, and interference factors need further confirmation.
Methods: Blood with heparin-lithium or EDTA-K2 were collected, stability of epinephrine (EPI), norepinephrine (NE), dopamine (DA), metanephrine (MN), normetanephrine (NMN), 3-methoxytyramine (3-MT) in whole blood and plasma at room temperature and 4 °C for different storage times, stability of plasma MN, NMN and 3-MT at -20 °C and -80 °C were investigated.