A sensor system based on a piezoelectric resonator with a lateral electric field in the frequency range 6-7 MHz of the electric field for virus detection is described. Through use of the transmissible virus causing gastroenteritis in pigs and specific antibodies, the possibility of detecting the virus in suspension in real time was determined. It was found that the frequency dependence of the real and imaginary parts of the electrical impedance of such a resonator loaded with a virus suspension changes significantly after the addition of specific antibodies to the suspension. No changes are observed if the antibodies are not specific. Thus, the results obtained illustrate the possibility of detecting viruses in situ, directly in the liquid phase, if the change in the real or imaginary parts of the electrical impedance after the addition of antibodies is used as an analytical signal. The possibility of virus detection in the presence of foreign viral particles has been illustrated.