Purpose: The purpose of this study is to evaluate the dosimetric impact of Hounsfield unit (HU) variations in kilovoltage cone-beam computed tomography (kV-CBCT) based 3D dose calculation accuracy in the treatment planning system and its validation using measured treatment delivery dose (MTDD) derived dose metrics for Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) plans in Head and Neck (HN) Cancer.

Methods: CBCT dose calculation accuracy was evaluated for 8 VMAT plans on inhomogeneous phantom and 40 VMAT and IMRT plans of HN Cancer patients and validated using ArcCHECK diode array MTDD derived 3D dose metric on CT and CBCT.

Results: The mean percentage dose difference between CBCT and CT in TPS (ΔD(CBCT-CT)) and 3DVH (ΔD(CBCT-CT)) were compared for the corresponding evaluation dose metrics (D, D, D, D, D, D, D, D) of all PTVs and OARs in phantom and patients. ΔD(CBCT-CT) and ΔD(CBCT-CT) for all evaluation dose points of all PTVs and OARs were less than 2.55% in phantom and 2.4% in HN patients. The Pearson correlation coefficient (r) between ΔD(CBCT-CT) and ΔD(CBCT-CT) for all dose points in all PTVs and OARs showed a strong to moderate correlation in phantom and patients with p < 0.001.

Conclusions: This study evaluated and validated the potential feasibility of kV-CBCT for treatment plan 3D dose reconstruction in clinical decision making for Adaptive radiotherapy on CT in Head and Neck cancer.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2022.02.014DOI Listing

Publication Analysis

Top Keywords

dose calculation
12
calculation accuracy
12
derived dose
12
dose metrics
12
ptvs oars
12
dose
11
kv-cbct based
8
based dose
8
head neck
8
mtdd derived
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!