Objective: To monitor ototoxicity, air conduction (AC) extended high frequency (EHF) thresholds can be measured up to 16 kHz. However, conductive hearing loss might influence these results. This is unfortunate because the EHF thresholds are important to follow the impact of ototoxic medication during therapy. Therefore a suitable bone conduction (BC) transducer and norm values for EHF BC measurements are needed.

Design: In this study three different BC transducers were used: the B71 (Radioear), the KH70 (Präcitronic), and the KLH96 (Westra). Hearing thresholds were measured from 0.125 to 16 kHz using AC transducers (Telephonics TDH39, Sennheiser HDA200), and BC thresholds from 0.25 to 8 kHz with the B71, and from 0.25 to 16 kHz with the KLH96 and KH70.

Study Sample: 60 ears of 30 normal hearing subjects were measured.

Results: The KLH96 showed the highest output for the high frequencies, and distortion measurements were similar to the KH70. The results show that EHF measurements are possible using the KLH96 and KH70 bone conductors.

Conclusion: EHF BC measurements are reliable when using the KLH96 and KH70 bone conductors. The extended force sensitivity of the used artificial mastoid should be determined for a proper EHF BC calibration.

Download full-text PDF

Source
http://dx.doi.org/10.1080/14992027.2022.2034059DOI Listing

Publication Analysis

Top Keywords

ehf measurements
12
extended high-frequency
8
bone conduction
8
ehf thresholds
8
thresholds measured
8
klh96 kh70
8
kh70 bone
8
ehf
6
bone
5
klh96
5

Similar Publications

: Conventional audiometry assesses hearing in the frequency range of 250-8000 Hz, while humans can detect sounds from 20 to 20,000 Hz. Hearing sensitivity above 8000 Hz can be assessed through "Extended high-frequency audiometry (EHFA)", which is useful for early detection and monitoring of hearing loss and counseling for tinnitus patients. Despite its usefulness, age significantly affects hearing sensitivity.

View Article and Find Full Text PDF

Objectives: Studies on the relationship between occupational noise exposure and extended high-frequency (EHF) hearing loss are limited. This study investigated the relationship between occupational noise exposure and EHF hearing loss in workers exposed to noise as measured by sound pressure level, exposure duration, and kurtosis to help provide a basis for early detection and prevention of hearing loss in noise-exposed workers.

Design: A cross-sectional survey was conducted among 602 participants with 472 noise-exposed workers and 130 non-noise-exposed controls.

View Article and Find Full Text PDF

Objectives: Valid wireless automated Békésy-like audiometry (ABA) outside a sound booth that includes extended high frequencies (EHF) would increase access to monitoring programs for individuals at risk for hearing loss, particularly those at risk for ototoxicity. The purpose of the study was to compare thresholds obtained with (1) manual audiometry using an Interacoustics Equinox and modified Hughson-Westlake 5 dB threshold technique to automated audiometry using the Wireless Automated Hearing Test System (WAHTS) and a Békésy-like 2 dB threshold technique inside a sound booth, and (2) ABA measured in the sound booth to ABA measured outside the sound booth.

Design: Cross-sectional study including 28 typically developing children and adolescents (mean = 14.

View Article and Find Full Text PDF

Hearing loss, particularly age-related hearing loss, significantly impacts health and quality of life worldwide. While much of the research has focused on older adults, the early stages of hearing decline remain relatively unexplored. Longitudinal studies examining hearing changes across the adult lifespan, especially at extended high frequencies (EHFs), are scarce.

View Article and Find Full Text PDF

Rationale: Sleep-related hypoventilation disorder (SHD) is common among obese patients with obstructive sleep apnea (OSA), but the pathological endotypes associated with obesity and SHD remain unclear.

Objectives: To investigate relationship between endotypes with body mass index (BMI) among patients with OSA, and to explore endotypic traits of patients with comorbid SHD.

Methods: We prospectively collected polysomnographic studies of 1364 patients with OSA, and overnight transcutaneous CO2 (TcCO2) measurements among 420 obese patients.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!