Experimental evaluation of the performance of large bore aspiration catheters.

J Neuroradiol

Division of Diagnostic and Interventional Neuroradiology, Geneva University Hospitals, 4 rue Gabrielle-Perret-Gentil, Geneva CH-1205, Switzerland. Electronic address:

Published: February 2023

Background And Purpose: Mechanical thrombectomy using an aspiration catheter (AC) is widely performed in patients with acute ischemic stroke due to large vessel occlusion. AC diameter directly impacts aspiration performance, which has led device companies to develop large bore ACs. The purpose of this study was to evaluate the in vitro performance of several commercially available large bore ACs.

Materials And Methods: We conceived an experimental set up to assess tracking force, aspiration flow rate, and distal end tensile force pre- and post-thrombectomy of ACs including Sofia 6 (Microvention Inc., California, USA), React 71 (Medtronic Neurovascular, Irvine, California, USA), Jet 7 (Penumbra Inc, Alameda, California, USA), Catalyst 7 (Stryker Neurovascular, Kalamazoo, Michigan, USA) and Embovac 071 (Johnson & Johnson Medical Ltd, UK).

Results: React 71 and Sofia 6 had significantly lower trackability force compared to the other ACs tested. Distal AC segment tensile force was highest for React 71 and lowest for Jet 7. Jet 7 had the highest internal diameter and the highest aspiration flow rate.

Conclusion: The data from this comparative analysis may assist clinicians in selecting the appropriate AC and highlights the need for key performance criteria during the development of next-generation large-bore ACs for optimal AC performance.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.neurad.2022.02.007DOI Listing

Publication Analysis

Top Keywords

large bore
12
california usa
12
aspiration flow
8
tensile force
8
performance
5
aspiration
5
experimental evaluation
4
evaluation performance
4
large
4
performance large
4

Similar Publications

Large-bore aspiration thrombectomy for the treatment of pulmonary embolism in octogenarians.

CVIR Endovasc

January 2025

Department of Radiology and Imaging Sciences, Indiana University School of Medicine, 550 University Blvd, Room 0641, Indianapolis, IN, 46202, USA.

Purpose: To evaluate outcomes in patients aged ≥ 80 years following large-bore aspiration thrombectomy (LBAT) for the treatment of pulmonary embolism (PE).

Materials And Methods: All patients ≥ 80 years of age with PE treated via LBAT at a single center were analyzed from September 2019 - August 2024. This included the octogenarian subgroup from a recently published retrospective analysis assessing all PE patients treated with LBAT at our center between September 2019 and January 2023.

View Article and Find Full Text PDF

Although venous stent placement is an established treatment for patients with deep vein thrombosis (DVT) and post-thrombotic syndrome (PTS), stent patency rate may be as low as 80% at 12 months. Treatment of stent occlusion requires the removal of a large amount of fresh or organized thrombus from stents with a diameter of 12-20 mm. The feasibility of large-bore rotational thrombectomy in combination with angioplasty to treat thrombosed venous iliofemoral or inferior vena cava stents has not been investigated yet.

View Article and Find Full Text PDF

Transcatheter mass extraction of left-sided cardiac masses has gained popularity in recent years, with scarce data on effectiveness, safety, and types of devices used. Mostly, left-sided aspirations use mechanical and continuous-flow-mediated devices (the AngioVac [AngioDynamics] and Penumbra systems [Penumbra]). To our knowledge, the use of manual aspiration devices has not been reported yet.

View Article and Find Full Text PDF

We report the case of a previously independent 82-year-old female who experienced acute hemodynamic and respiratory deterioration requiring inotropic support due to a fat embolism during revision hip arthroplasty. Computed tomography pulmonary angiography demonstrated fat embolism, and transesophageal echocardiogram showed evidence of right ventricle strain and fat embolism in-transit in the right heart, as well as a moderate patent foramen ovale. Under transesophageal echocardiogram and intravascular ultrasound guidance, the Inari FlowTriever thrombectomy device was used successfully to retrieve the fat embolism with immediate hemodynamic improvement, no complications, and uneventful recovery.

View Article and Find Full Text PDF

Safety of accessing brachial veins for large-bore upper extremity venous thrombectomy using ClotTriever Thrombectomy System.

CVIR Endovasc

January 2025

Department of Radiology, Section of Vascular and Interventional Radiology, University of Washington, 1959 Northeast Pacific Street, Seattle, WA, 98195, USA.

Purpose: To evaluate access site adverse events following ClotTriever-mediated large-bore mechanical thrombectomy via small upper extremity deep veins (< 6-mm).

Materials And Methods: Twenty patients, including 24 upper extremity venous access sites, underwent ClotTriever-mediated large-bore thrombectomy of the upper extremity and thoracic central veins for symptomatic deep vein obstruction unresponsive to anticoagulation. Patients without follow-up venous duplex examinations (n = 3) were excluded.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!