Background: Traditional viscoelastic clotting tests are significantly impacted by the operator and environmental variation. The VCMVet coagulation monitor could provide a more user-friendly alternative for veterinary practices.

Objectives: We aimed to determine if environmental vibration commonly encountered in veterinary practice alters the results of a point-of-care viscoelastic device, the VCMVet.

Methods: Nine fresh whole blood samples from healthy dogs were evaluated simultaneously using VCMVet instruments under four environmental conditions: (normal) alone and undisturbed on a raised tabletop, (centrifuge) on a countertop 6 inches from a centrifuge that operated at 12 000 rpm for 10 minutes every 20 minutes, (workspace) on a tabletop workspace in proximity to two heavy-use computers, and (gurney) on a rolling gurney at a walking pace for 10 minutes every 20 minutes. Results were compared between conditions using a Friedman test, and if this was significant (P-value < .05), it was followed by a Wilcoxon test for paired samples.

Results: Analysis of samples on a rolling gurney created obvious movement artifacts, and this condition was excluded from statistical analysis. The centrifuge condition resulted in a significantly higher alpha angle (median 49 degrees, interquartile range 4) than the normal condition (median 46 degrees, interquartile range 5, P = .0078). Other viscoelastic parameters were not significantly different between the normal, centrifuge, and workspace conditions.

Conclusions: The VCMVet is suitable for use in a busy veterinary environment but should be protected from vibration. The instrument does not produce reliable results when operated on a moving gurney, and it should be stationary during sample analysis.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vcp.13041DOI Listing

Publication Analysis

Top Keywords

coagulation monitor
8
minutes minutes
8
rolling gurney
8
median degrees
8
degrees interquartile
8
interquartile range
8
response vcmvet
4
viscoelastic
4
vcmvet viscoelastic
4
viscoelastic coagulation
4

Similar Publications

Background: Blood loss during liver transplantation (LT) remains a major concern associated with increased morbidity and reduced patient and graft survival. The high complexity of the procedure associated with the multifaceted origin of the bleeding urges early identification of high-risk patients and proper monitoring of hemostasis disorders in order to improve results. The accuracy of international normalized ratio (INR) and activated partial thromboplastin time (aPTT) to evaluate coagulation status in cirrhotic patients has been doubted.

View Article and Find Full Text PDF

Applicability of computed tomography and rhinoscopy in the diagnosis and monitoring of the treatment of epistaxis in a dog.

Braz J Vet Med

January 2025

Veterinarian, DSc. Departamento de medicina e cirurgia veterinária, Instituto de Veterinária, Universidade Federal Rural do Rio de Janeiro. Seropédica, RJ, Brazil.

Epistaxis is defined as bleeding from the nasal cavity and can be related to systemic causes leading to coagulation disorders, most commonly hemoparasitosis, or to localized changes in the nasal cavity itself (e.g., intranasal neoplasms).

View Article and Find Full Text PDF

Gene therapy in hemophilia: the dawn of a new era.

Res Pract Thromb Haemost

January 2025

Dipartimento di Fisiopatologia Medico-chirurgica e dei Trapianti, Università degli Studi di Milano, Milano, Italia.

Hemophilia A and B are hereditary bleeding disorders associated with the X chromosome, stemming from genetic defects in the coding of coagulation factor (F)VIII or FIX protein, leading to partial or complete deficiency. In the absence of effective prophylaxis, these deficiencies can result in irreversible joint damage, known as hemophilic arthropathy, and subsequent disability. Despite advancements in hemophilia treatment, individuals with severe forms of the disease continue to face a high risk of bleeding, particularly in instances of trauma or major surgical procedures.

View Article and Find Full Text PDF

Rationale: Warfarin is the most commonly used drug in patients with mechanical valve replacement. Acute liver damage after warfarin is rare but potentially harmful. We present a case of warfarin-induced gastrointestinal bleeding with liver injury, pharmacy monitoring, and its therapy.

View Article and Find Full Text PDF

Widely used second-generation anticoagulant rodenticides like brodifacoum are classified as persistent, bioaccumulative, and toxic. Widespread exposure of terrestrial and avian non-target species is well-known and recently hepatic anticoagulant rodenticide residues have been detected in wild fish. However, no sufficient data exist to interpret the effects of these findings on fish health.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!