Background: ABO antibody titres are important in many clinical decisions; however, much variability is observed in titre results. For reliable and reproducible titre results, automated ABO titration methods have been developed. In this 10-site study, we evaluated the equivalency of the automated ABO titration assays on the Galileo NEO, a fully automated blood bank analyzer (Immucor, Inc.) to manual titration with gel Column Agglutination Technology (CAT), as well as the reproducibility of both methods.

Materials And Methods: Ten different locations participated in this study. The equivalency study included 70 random samples at each site. The reproducibility study tested the same blinded 30-sample panel at each study site. Anti-A and anti-B IgM and IgG antibody titres were tested with both the automated and manual methods; additionally, dithiothreitol (DTT) treatment was used to inactivate IgM antibodies in the manual CAT method.

Results: The equivalency between CAT manual method and Galileo NEO automated titres at each site ranged from 38 to 88%; equivalency for each isotype was 66.2% for IgM, 60.6% for IgG, and 88.5% for DTT-treated IgG. The reproducibility study evaluated the titre variation of each sample obtained from the 10 sites. The average titre ranges (in doubling dilutions) for the automated and manual methods, respectively, were 2.15±1.0 and 4.03±1.8 for IgM, and 1.53±0.7 and 4.10±1.9 for IgG; for the manual DTT-treated IgG, the average titre range was 3.45±1.8 doubling dilutions.

Discussion: The results demonstrated that the Galileo NEO automated and manual CAT ABO titres are not equivalent. However, the study also demonstrated that titre reproducibility is enhanced with the Galileo NEO automated ABO titration assays relative to the manual CAT ABO titration method. Therefore, to improve management of patients receiving care across multiple institutions, our study supports the use of automated ABO titration.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9256507PMC
http://dx.doi.org/10.2450/2022.0197-21DOI Listing

Publication Analysis

Top Keywords

abo titration
20
automated abo
16
galileo neo
16
antibody titres
12
automated manual
12
manual cat
12
neo automated
12
automated
10
study
9
manual
9

Similar Publications

Background: The popularity of Low-Titer O Whole Blood (LTOWB) for treating trauma patients requires that donor centers and transfusion services make decisions on what titer testing capabilities to institute and an appropriate titer level threshold. This study compared the titer results determined by four methods to find a rate of agreement.

Study Design And Methods: Isohemagglutinin titers were tested on 300 plasma samples utilizing various methods, each determining IgM antibody levels by direct hemagglutination with A1 and B reference cells.

View Article and Find Full Text PDF

Background And Objectives: The storage temperature of immunohaematological reagents generally ranges from 2 to 8°C, and they should be utilised at room temperature. This study aimed to analyse the stability of immunohaematological reagents used in ABO and RhD typing.

Methods: The evaluation encompassed the potency, specificity, and integrity of anti-A, anti-B, anti-D, RhD control sera, and A and B red blood cells (RBC) reagents after long (8 h) and short (4 h) daily periods of exposure to room temperature (20-24°C), 5 days a week for 4 weeks.

View Article and Find Full Text PDF

Different roles of the heterodimer architecture of galectin-4 in selective recognition of oligosaccharides and lipopolysaccharides having ABH antigens.

J Biol Chem

August 2024

CIC bioGUNE, Bizkaia Technology Park, Derio, Bizkaia, Spain; Ikerbasque, Basque Foundation for Science, Bilbao, Bizkaia, Spain. Electronic address:

The dimeric architecture of tandem-repeat type galectins, such as galectin-4 (Gal-4), modulates their biological activities, although the underlying molecular mechanisms have remained elusive. Emerging evidence show that tandem-repeat galectins play an important role in innate immunity by recognizing carbohydrate antigens present on the surface of certain pathogens, which very often mimic the structures of the human self-glycan antigens. Herein, we have analyzed the binding preferences of the C-domain of Gal-4 (Gal-4C) toward the ABH-carbohydrate histo-blood antigens with different core presentations and their recognition features have been rationalized by using a combined experimental approach including NMR, solid-phase and hemagglutination assays, and molecular modeling.

View Article and Find Full Text PDF

Background And Objectives: Hemolytic transfusion reactions (HTRs) pose significant risks in transfused patients, with anti-A and anti-B antibodies in donor plasma being potential contributing factors. Despite advancements in component preparation, HTRs remain a concern, particularly with apheresis-derived platelets. This study aimed to determine the prevalence of high anti-A and anti-B titers among A, B, and O blood group donors and to explore factors associated with high titers.

View Article and Find Full Text PDF

Isohemagglutinin titration in pooled and apheresis platelets.

Transfusion

July 2024

Department of Laboratory Medicine and Pathology, University of Alberta, Edmonton, Alberta, Canada.

Background: Platelet inventory constraints necessitate ABO-incompatible platelet transfusion. Many minimize the hemolytic impact by confirming low titre (LT) donor isohemagglutinins. This process is costly.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!