Aim And Objective: To compare the results of a new automated glaucoma test-Pattern-Noise (PANO)-to the Humphrey Visual Field Analyzer-II (HFA), the fundus area cup-to-disk ratio (CDR), and a frequency doubling technology (FDT) stimulus.

Materials And Methods: This was a prospective study performed in the West-Region of Cameroon. Two hundred and nineteen eyes of 122 adult patients were included with a clinical suspicion of normal-tension or primary open-angle glaucoma and no other major ocular pathology. Eyes were examined with PANO, HFA (24-2 SITA standard), and FDT-stimulus in a randomized order followed by clinical assessment of the CDR.

Results: Parametric correlation of the mean contrast threshold of PANO with the mean contrast threshold of FDT-stimulus, total deviation of HFA, and area CDR was 0.94, -0.85, and 0.62, respectively ( < 0.001 for all values). Spatial distribution of sensitivity thresholds is highly correlated ( < 0.001) at all points in the visual field between PANO and HFA. With cut-off values of 3 ± 1 dB for HFA mean deviation and 4 ± 1 for PANO mean contrast threshold and after eliminating borderline cases, PANO's sensitivity was 95% and specificity 60%. The mean patient age was 45.2 ± 15.8 years. Mean thresholds of PANO and FDT-stimulus decreased with increasing age. Mean examination time was 7.1 ± 1.8 minutes for PANO, 5.9 ± 1.3 minutes for HFA, and 4.7 ± 1.3 minutes for FDT-stimulus. The mean percentage of false-positives per examination was 4.95% for PANO, 4.62% ( = 0.025) for FDT-stimulus, and 2.10% for HFA.

Conclusion: The results showed that PANO was successful in suspecting the presence of glaucoma. Pattern-Noise examination led to findings that were significantly correlated to HFA, FDT stimulus, and area CDR. Some patterns of defect were also correlated. Furthermore, PANO showed a reasonable examination time and error rate.

Clinical Significance: Affordable and robust visual field devices are lacking in large parts of the developing world. Comparing them to established methods is a prerequisite to their clinical use.

How To Cite This Article: Hannen T, El-Khoury S, Patel R, Comparison of the Automated Pattern-Noise (PANO) Glaucoma Test with the HFA, an FDT Stimulus, and the Fundus Area Cup-to-disk Ratio. J Curr Glaucoma Pract 2021;15(3):132-138.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8807941PMC
http://dx.doi.org/10.5005/jp-journals-10078-1317DOI Listing

Publication Analysis

Top Keywords

hfa fdt
12
fdt stimulus
12
fundus area
12
area cup-to-disk
12
cup-to-disk ratio
12
visual field
12
contrast threshold
12
pano
11
hfa
9
comparison automated
8

Similar Publications

Diagnostic Accuracy of Frequency-Doubling Technology and the Moorfields Motion Displacement Test for Glaucoma.

Ophthalmol Glaucoma

May 2023

Francis I. Proctor Foundation, University of California, San Francisco, California; Department of Ophthalmology, University of California, San Francisco, California. Electronic address:

Purpose: Portable perimetric testing could be useful for community-based glaucoma screening programs. Frequency-doubling technology (FDT) and the Moorfields motion displacement test (MDT) are portable perimeters that have shown promise as potential screening tools for glaucoma. This study's goal was to determine the diagnostic accuracy of FDT and MDT for visual field defects and glaucoma.

View Article and Find Full Text PDF

Aim And Objective: To compare the results of a new automated glaucoma test-Pattern-Noise (PANO)-to the Humphrey Visual Field Analyzer-II (HFA), the fundus area cup-to-disk ratio (CDR), and a frequency doubling technology (FDT) stimulus.

Materials And Methods: This was a prospective study performed in the West-Region of Cameroon. Two hundred and nineteen eyes of 122 adult patients were included with a clinical suspicion of normal-tension or primary open-angle glaucoma and no other major ocular pathology.

View Article and Find Full Text PDF

Purpose: Describing the psychometric characteristics and diagnostic accuracy of the Accelerator 4-Alternative Forced-Choice Flicker Test prototype (A4FTp) for detecting chronic open angle glaucoma (COAG).

Methods: A4FTp measures temporally-modulated flicker thresholds in regions of the visual field with high susceptibility to glaucomatous loss. We initially evaluated its psychometric properties on 20 normals (aged 33.

View Article and Find Full Text PDF

Purpose: The Humphrey Matrix (FDT2) may be more sensitive in detecting glaucomatous visual field loss than SITA standard automated perimetry (SAP) performed on the Humphrey Field Analyzer (HFA). Therefore, FDT may be a good candidate to determine disease progression in patients with glaucoma. Our aim was to test the hypothesis that automated perimetry using the FDT2 would be equal to, or more effective than, HFA SITA-Standard, in identifying glaucomatous progression.

View Article and Find Full Text PDF

Aims: The aim of this study was to compare the visual field test results in healthy children obtained via the Humphrey matrix 24-2 threshold program and standard automated perimetry (SAP) using the Swedish interactive threshold algorithm (SITA)-Standard 24-2 test.

Materials And Methods: This prospective study included 55 healthy children without ocular or systemic disorders who underwent both SAP and frequency doubling technology (FDT) perimetry visual field testing. Visual field test reliability indices, test duration, global indices (mean deviation [MD], and pattern standard deviation [PSD]) were compared between the 2 tests using the Wilcoxon signed-rank test and paired t-test.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!