Background: The Aptima Combo 2 (AC2) assay manufactured by Hologic, Inc., detects Neisseria gonorrhoeae and/or Chlamydia trachomatis (CT) in urogenital and extragenital specimens by targeting either a 16S rRNA (N. gonorrhoeae) or 23S rRNA (CT) region. In 2019, a mutation (C1515T) in the 23S rRNA region was reported to cause false-negative/equivocal results in specimens collected in Finland. Specimens containing this variant (Fl-nvCT) were also discovered internationally. Working with specimens submitted to a large commercial laboratory, we sought to determine if this variant was also present in the United States.
Methods: A subset (n = 401) of specimens tested with the AC2 assay collected during a 5-week period in late 2019/early 2020 were evaluated using an updated AC2 assay.
Results: Although the FI-nvCT variant was not detected within this specimen panel, 2 CT variants containing 23S rRNA mutations (A1518G, G1526A) were identified. The updated AC2 assay targeting an additional region of the 23S rRNA detected both of these variants. A retrospective study of >18 million AC2 results tested between 2018 and 2019 did not display a decrease in CT positivity.
Conclusions: Although we did not detect the Fl-nvCT variant among US specimens, we show evidence that the low occurrence of similar diagnostic-escape mutants can be detected with an updated AC2 assay using multiple 23S rRNA targets.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9133125 | PMC |
http://dx.doi.org/10.1097/OLQ.0000000000001617 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!