Impact of metal artefacts on subjective perception of image quality of 13 CBCT devices.

Clin Oral Investig

OMFS IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, KU Leuven and Oral and Maxillofacial Surgery, University Hospitals Leuven, Leuven, Belgium.

Published: June 2022

Objectives: The overall objective of this study was to assess how metal artefacts impact image quality of 13 CBCT devices. As a secondary objective, the influence of scanning protocols and field of view on CBCT image quality with and without metal artefacts was also assessed.

Materials And Methods: CBCT images were acquired of a dry human skull phantom considering three clinical simulated conditions: one without metal and two with metallic materials (metallic pin and implant). An industrial micro-CT was used as a reference to register the CBCT images. Afterwards, four observers evaluated 306 representative image slices from 13 devices, ranking them from best to worst. Furthermore, within each device, medium FOV and small FOV standard images were compared. General linear mixed models were used to assess subjective perception of examiners on overall image quality in the absence and presence of metal-related artefacts (p < 0.05).

Results: Image quality perception significantly differed amongst CBCT devices (p < 0.05). Some devices performed significantly better, independently of scanning protocol and clinical condition. In the presence of metal artefacts, medium FOV standard scanning protocols scored significantly better, while in the absence of metal, small FOV standard yielded the highest performance.

Conclusions: Subjective image quality differs significantly amongst CBCT devices and scanning protocols. Metal-related artefacts may highly impact image quality, with a significant device-dependent variability and only few scanners being more robust against metal artefacts. Often, metal artefact expression may be somewhat reduced by proper protocol selection.

Clinical Relevance: Metallic objects may severely impact image quality in several CBCT devices.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00784-022-04409-wDOI Listing

Publication Analysis

Top Keywords

image quality
16
metal artefacts
12
subjective perception
8
quality cbct
8
cbct devices
8
cbct images
8
image
5
cbct
5
impact metal
4
artefacts
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!