Antimicrobial susceptibility testing determined by Alfred 60/AST (Alifax®) in a multi-sites lab: performance's evaluation and optimization of workflow.

J Microbiol Methods

Microbiology Department, Laboratoire Hospitalier Universitaire de Bruxelles - Universitair Laboratorium Brussel (LHUB-ULB), Université Libre de Bruxelles, Brussels, Belgium; Medicine and Pharmacy departments, University of Mons (UMONS), Mons, Belgium.

Published: March 2022

Purpose: New techniques are needed to speed-up the identification and antimicrobial susceptibility testing (AST) of bacteria associated with bloodstream infections. Alfred 60/AST (Alifax®, Polverara, Italy) performs AST by light scattering directly from positive blood cultures.

Methods: We evaluated Alfred 60/AST performances for 4 months. Each new episode of bacteraemia was included and AST were compared to either our rapid automated AST (Vitek® 2) or disk diffusion method. The discrepancies were investigated using Etest®. The time-to-result (TTR) was evaluated by comparing the blood volume inserted into Alfred 60/AST, i.e. 2 versus 7 blood drops. Taking into account the TTR, the workflow of positive blood cultures and the availability of AST results was studied in order to optimize the implementation of Alfred 60/AST.

Results: A total of 249 samples and 1108 antibiotics for AST were tested. After exclusion of unavailable results, 1008 antibiotics were analysed. 94.9% (n = 957/1008) of the antibiotics showed categorical agreement. There were 14 very major errors (VME), 24 major errors (ME) and 13 minor errors (mE). The VME were mostly related to clindamycin (64.3%) whereas meropenem and piperacillin-tazobactam constituted the major part (37.5% and 61.5%) of ME and mE respectively. Results were highly reliable for Enterobacterales and enterococci. The mean TTR ranged between 4.3 and 6.3 h and was statistically 20 min faster when applying the 7 blood drops protocol. We showed that Alfred 60/AST could give relievable results within working hours for positive blood culture which are flagged the same day between 12:00 am and 12:00 pm.

Conclusion: Our study confirmed that Alfred 60/AST gives reliable AST results in a short period of time, especially for Enterobacterales and enterococci. AST could thus be easily obtained the same day of a positive blood culture. Clinical impact studies are mandatory to validate a 24/24 working.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.mimet.2022.106433DOI Listing

Publication Analysis

Top Keywords

alfred 60/ast
24
positive blood
16
antimicrobial susceptibility
8
susceptibility testing
8
60/ast alifax®
8
ast
8
blood drops
8
major errors
8
errors vme
8
enterobacterales enterococci
8

Similar Publications

affects elderly people and is associated with individuals with urological-related predispositions, but can be found in a variety of locations, such as cutaneous, intraabdominal, genitourinary and surgical infections. Disseminated infections occur less frequently and are by and large related to urinary tract colonisation. This pathogen is often neglected due to growth requirements, especially in urinary tract infections.

View Article and Find Full Text PDF

Antimicrobial susceptibility testing determined by Alfred 60/AST (Alifax®) in a multi-sites lab: performance's evaluation and optimization of workflow.

J Microbiol Methods

March 2022

Microbiology Department, Laboratoire Hospitalier Universitaire de Bruxelles - Universitair Laboratorium Brussel (LHUB-ULB), Université Libre de Bruxelles, Brussels, Belgium; Medicine and Pharmacy departments, University of Mons (UMONS), Mons, Belgium.

Purpose: New techniques are needed to speed-up the identification and antimicrobial susceptibility testing (AST) of bacteria associated with bloodstream infections. Alfred 60/AST (Alifax®, Polverara, Italy) performs AST by light scattering directly from positive blood cultures.

Methods: We evaluated Alfred 60/AST performances for 4 months.

View Article and Find Full Text PDF

The aim of this study was to evaluate the performance of the new automated system AlfredAST which is based on light scattering technology for rapid susceptibility testing directly from positive blood cultures as well as its applicability in the routine laboratory workflow. We evaluated 176 significant episodes of bacteremia due to 92 Gram-negative and 84 Gram-positive bacteria. The antimicrobial agents tested were ceftriaxone, ciprofloxacin, gentamicin, meropenem, piperacillin-tazobactam, and colistin for Gram negatives and cefoxitin, vancomycin, linezolid, and daptomycin for Gram positives.

View Article and Find Full Text PDF

Background: Blood stream infections (BSIs) are a major cause of morbidity and mortality. The time from taking blood cultures to obtain results of antibiotic sensitivity can be up to five days which impacts patient care. The Alfred 60 AST™ can reduce laboratory time from positive culture bottle to susceptibility results from 16 to 25 h to 5-6 h, transforming patient care.

View Article and Find Full Text PDF

Background: A bacteremia diagnosis with speeded-up identification and antimicrobial susceptibility testing (AST) is mandatory to adjust empirical broad-spectrum antibiotherapy and avoid the emergence of multi-resistant bacteria. Alfred 60 (Alifax, Polverara, PD, Italy) is an innovative automated system based on light scattering measurements allowing direct AST from positive blood cultures with rapid results. In this study we aimed to evaluate the system's performances and turnaround time (TAT) compared to routine AST.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!