Objective: To study the effectiveness of a new spinal protection device for preventing and treating osteoporotic vertebral compression fractures (OVCFs) by finite element analysis (FEA).
Methods: One healthy volunteer and one patient with 1-segment lumbar vertebral compression fractures were included in this experimental study. The DICOM files of two different lumbar spiral computed tomography (CT) scans were converted into STL files, and 3D finite element models of the lumbar spine were generated for normal and L1 vertebral fracture spines. A new type of spinal protection device was applied to reduce the stress on the anterior vertebral edge and direct the center of gravity posteriorly. The stress distribution characteristics of different finite element models of the lumbar spine were analyzed, revealing the characteristics of the stress distributed along the spine under the action of the new spinal protection device.
Results: Under normal conditions, the stress was mainly distributed in the middle and posterior columns of the spine. When the anterior border of the L1 vertebral body was fractured and collapsed, the stress distribution shifted toward the anterior column due to the center of gravity being directed forward. According to finite element analysis of the spine with the new protection device, the stress in the middle and posterior columns tended to increase, and that in the anterior column decreased. After the new type of spinal fixation device was applied, the stress at the L1 and L2 vertebral endplates decreased to a certain extent, especially that at the L1 vertebral body. The maximum stress on the L1 vertebral body decreased by 20% after the auxiliary device was applied.
Conclusions: According to the FEA results, the new spinal protection device can effectively prevent and treat osteoporotic vertebral compression fractures (OVCFs), and can alter the stress distribution in the spine and reduce the stress in the anterior column of the vertebral body, especially in vertebral compression fractures.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8926982 | PMC |
http://dx.doi.org/10.1111/os.13220 | DOI Listing |
World J Orthop
January 2025
Undip Biomechanics Engineering and Research Centre, Universitas Diponegoro, Semarang 50275, Central Java, Indonesia.
Total hip arthroplasty for adults with sequelae from childhood hip disorders poses significant challenges due to altered anatomy. The paper published by Oommen reviews the essential management strategies for these complex cases. This article explores the integration of finite element analysis (FEA) to enhance surgical precision and outcomes.
View Article and Find Full Text PDFKorean J Orthod
January 2025
Department of Orthodontics, Hacettepe University, Ankara, Türkiye.
Objective: This study aimed to evaluate the effects of attachment design on maxillary molar distalization and simultaneous expansion during distalization, and the influence of third molars on distalization in the clear aligner technique using the finite element method.
Methods: Six models were created to evaluate three different attachment designs on the second molars. Model I: employed a vertical rectangular attachment; Model II: used a vertical rectangular attachment with the presence of the third molar; Model III: used a combined semi-elliptical attachment; and Model IV: featured an opposed semi-elliptical attachment with buccal and palatal components.
Korean J Orthod
January 2025
Private Practice, Ankara, Türkiye.
Objective: The effect of different attachment positions on torque control during the labialization of maxillary lateral incisors with clear aligners was evaluated using finite element analysis.
Methods: Anatomical data acquired through cone-beam computed tomography, combined with the design of 0.625-mm-thick aligners and horizontal attachments, were integrated into the software.
J Orthop Surg Res
January 2025
Department of Orthopaedics and Traumatology, Selçuk University Faculty of Medicine, Konya, 42130, Türkiye.
Background And Objectives: Bennett fractures are intra-articular fractures of the first metacarpal, typically resulting from axial loading. The optimal surgical method for fixation remains a topic of ongoing debate. This study aims to evaluate and compare the biomechanical properties of different fixation techniques.
View Article and Find Full Text PDFSci Rep
January 2025
School of Instrumental Science and Engineering, Harbin Institute of Technology, 150001, Harbin, People's Republic of China.
In this study, a method for predicting the thermal shock life of coatings is proposed, and a model for predicting the thermal shock life of coatings based on high temperature thermal shock life test and three-dimensional heat transfer analysis is established. Firstly, the thermal shock life of coatings at different cooling and heating cycle temperatures is obtained through a designed thermal shock life testing device for silicide coatings at a wide-temperature range from 500℃ to 3000℃. Secondly, the actual thickness of the coating and the continuous oxidation in the thermal shock life test are taken into consideration.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!