Objective: This study aimed to investigate the relationship between cochlear implant (CI) electrode distances to the cochlea's inner wall (the modiolus) and electrical impedance measurements made at the CI's electrode contacts. We introduced a protocol for "three-point impedances" in which we recorded bipolar impedances in response to monopolar stimulation at a neighboring electrode. We aimed to assess the usability of three-point impedances and two existing CI impedance measurement methods (monopolar and four-point impedances) for predicting electrode positioning during CI insertion.

Methods: Impedances were recorded during stepwise CI electrode array insertions in cadaveric human temporal bones. The positioning of the electrodes with respect to the modiolus was assessed at each step using cone beam computed tomography. Linear mixed regression analysis was performed to assess the relationship between the impedances and electrode-modiolar distances. The experimental results were compared to clinical impedance data and to an existing lumped-element model of an implanted CI.

Results: Three-point and four-point impedances strongly correlated with electrode-modiolar distance. In contrast, monopolar impedances were only minimally affected by changes in electrode positioning with respect to the modiolus. An overall model specificity of 62% was achieved when incorporating all impedance parameters. This specificity could be increased beyond 73% when prior expectations of electrode positioning were incorporated in the model.

Conclusion: Three-point and four-point impedances are promising measures to predict electrode-modiolar distance in real-time during CI insertion.

Significance: This work shows how electrical impedance measurements can be used to predict the CI's electrode positioning in a biologically realistic model.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2022.3150239DOI Listing

Publication Analysis

Top Keywords

electrode positioning
16
three-point four-point
12
impedance measurements
12
four-point impedances
12
electrode
9
cochlear implant
8
implant electrode
8
electrical impedance
8
ci's electrode
8
impedances
8

Similar Publications

The tunability of the energy bandgap in the near-infrared (NIR) range uniquely positions colloidal lead sulfide (PbS) quantum dots (QDs) as a versatile material to enhance the performance of existing perovskite and silicon solar cells in tandem architectures. The desired narrow bandgap (NBG) PbS QDs exhibit polar (111) and nonpolar (100) terminal facets, making effective surface passivation through ligand engineering highly challenging. Despite recent breakthroughs in surface ligand engineering, NBG PbS QDs suffer from uncontrolled agglomeration in solid films, leading to increased energy disorder and trap formation.

View Article and Find Full Text PDF

Electrochemical energy storage plays a critical role in the transition to clean energy. With the growing demand for efficient and sustainable energy solutions, supercapacitors have gained significant attention due to their high specific capacitance, rapid charge/discharge capabilities, long lifespan, safe operation across various temperatures, and minimal maintenance needs. This study introduces a novel approach for the synthesis of high-performance supercapacitor electrodes by using MnNi-MOF-74 as a precursor.

View Article and Find Full Text PDF

Refining computer-assisted SEEG planning with spatial priors - A novel comparison of implantation strategies across adult and paediatric centres.

Neurophysiol Clin

January 2025

Developmental Neurosciences, UCL Great Ormond Street Institute of Child Health, London, WC1N 1EH, UK; Department of Neurosurgery, Great Ormond Street Hospital for Children, London, WC1N 3JH, UK.

Objectives: Computer-assisted planning (CAP) allows faster SEEG planning and improves grey matter sampling, orthogonal drilling angles to the skull, reduces risk scores and minimises intracerebral electrode length. Incorporating prior SEEG trajectories enhances CAP planning, refining output with centre-specific practices. This study significantly expands on the previous work, compares priors libraries between two centres, and describes differences between SEEG in adults and children in these centres.

View Article and Find Full Text PDF

Background And Objectives: Surgical planning is critical to achieve optimal outcome in deep brain stimulation (DBS). The relationship between clinical outcomes and DBS electrode position relative to subthalamic nucleus (STN) is well investigated, but the role of surgical trajectory remains unclear. We sought to determine whether preoperatively planned DBS lead trajectory relates to adequate motor outcome in STN-DBS for Parkinson's disease (PD).

View Article and Find Full Text PDF

Objective: Electrical Impedance Tomography (EIT) is a non-invasive technique used for lung imaging. A significant challenge in EIT is reconstructing images of deeper thoracic regions due to the low sensitivity of boundary voltages to internal conductivity variations. The current injection pattern is decisive as it influences the current path, boundary voltages, and their sensitivity to tissue changes.

View Article and Find Full Text PDF