Ambulatory Heart Rate Variability Monitoring: Comparisons Between the Empatica E4 Wristband and Holter Electrocardiogram.

Psychosom Med

From the Durham Veterans Affairs Medical Center (Van Voorhees, P.A. Dennis, Patel, Calhoun, M.F. Dennis, Beckham); Department of Psychiatry and Behavioral Sciences, Duke University Medical Center (Van Voorhees, P.A. Dennis, Watkins, Patel, Calhoun, M.F. Dennis, Beckham); Veterans Affairs Mid-Atlantic Region Mental Illness Research, Education, and Clinical Center (Calhoun, Beckham), Durham; and Durham Veterans Affairs Center for Health Services Research in Primary Care (Calhoun), Durham, North Carolina.

Published: March 2022

Objective: Heart rate variability (HRV) is a useful index of psychological and physiological stress. Although several wristband devices have purported to measure HRV, none have demonstrated reliability when compared with the criterion-standard Holter monitor. We evaluated the reliability of HRV readings from the Empatica E4 wristband compared with a Holter monitor over a 24-hour period of simultaneous monitoring.

Methods: Agreement between the monitors was assessed by examining correlations and intraclass correlations (ICCs) for fixed sets in 13 individuals in a treatment trial for posttraumatic stress disorder (4 women; mean [standard deviation] age = 51.92 [6.17] years). Agreement was calculated at 1-second and 5-minute intervals for interbeat intervals (IBIs) and for 5-minute intervals of the root mean square of successive differences between normal heartbeats (RMSSD) and standard deviation of all normal R-R intervals (SDNN). Agreement across the entire 24-hour observation period was also measured. Frequency-domain measures of HRV could not be calculated because of too much missing data from the E4.

Results: Although high interdevice correlations and ICCs were observed between the E4 and Holter monitors for IBIs at 1-second (median r = 0.88; median ICC = 0.87) and 5-minute (median r = 0.94; median ICC = 0.94) intervals, reliabilities for 5-minute RMSSD (median r = -0.09; median ICC = -0.05) and 5-minute SDNN (median r = 0.48; median ICC = 0.47) were poor. Agreement between the devices on 24-hour measures of HRV was satisfactory (IBI: r = 0.97, ICC = 0.97; RMSSD: r = 0.77, IBI = 0.76; SDNN: r = 0.92, IBI = 0.89).

Conclusions: Findings suggest that the low reliability of Empatica E4 as compared with the Holter monitor does not justify its use in ambulatory research for the measurement of HRV over time periods of 5 minutes or less.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8851683PMC
http://dx.doi.org/10.1097/PSY.0000000000001010DOI Listing

Publication Analysis

Top Keywords

median icc
16
holter monitor
12
heart rate
8
rate variability
8
empatica wristband
8
compared holter
8
correlations iccs
8
5-minute intervals
8
measures hrv
8
median
8

Similar Publications

Objective: To test the reliability of the Outcome Measures in Rheumatology Giant cell arteritis (GCA) Ultrasonography Score (OGUS) and other composite scores in a patient-based exercise involving experts and non-experts in vascular ultrasonography.

Methods: Six GCA patients were scanned twice (two rounds separated ≥3 hours) by 12 experts and 12 non-experts. Non-experts received 90 min of theoretical and 240 min of practical training between rounds 1 and 2.

View Article and Find Full Text PDF

Background: Test-retest reliability for existing cognitive screening tests is typically poor and most have ceiling effects and restricted score ranges that mask the presence of subtle decline. The Brief Assessment of Cognitive Health (BACH) is a computerized cognitive screening tool that patients complete independently. It includes a complex memory test without ceiling effects and brief mood and history questions.

View Article and Find Full Text PDF
Article Synopsis
  • Human Aichi virus 1 (AiV-1) is a picornavirus linked to gastroenteritis and is found frequently in environmental waters, indicating potential fecal contamination.
  • Recent research examined 450 water samples from a Tunisian drinking water treatment plant and Sidi Salem dam, revealing 18.9% tested positive for AiV-1 with varying viral loads throughout different treatment stages.
  • The presence of infectious AiV-1 particles poses a public health risk, and the study highlights the effectiveness of the integrated cell culture approach combined with quantitative molecular detection (ICC-RT-qPCR) for monitoring viruses in water.
View Article and Find Full Text PDF

A decrease in REM time during polysomnography (PSG) in patients with obstructive sleep apnea (OSA) can result in underestimation of apnea hypopnea index (AHI). We propose adjusting AHI to normalized REM% in subjects with REM% ≤15% to avoid under diagnosis of OSA. All children who completed diagnostic PSG from 2016 to 2023 with REM% of ≤ 15% of total TST were selected for adjustment.

View Article and Find Full Text PDF

Validation of a novel tool for automated tooth modelling by fusion of CBCT-derived roots with the respective IOS-derived crowns.

J Dent

December 2024

OMFS-IMPATH Research Group, Department of Imaging and Pathology, Faculty of Medicine, KU Leuven, Leuven, Belgium; Department of Dental Medicine, Karolinska Institute, Stockholm, Sweden. Electronic address:

Article Synopsis
  • This study aimed to validate a novel AI tool that combines dental images from two different sources, cone beam computed tomography (CBCT) and intraoral scanners (IOS), to create accurate tooth models.
  • The evaluation involved analyzing a dataset from 30 patients, calculating various accuracy metrics for the AI-generated models, and comparing them to both expert manual methods and existing segmentation approaches.
  • Results indicated that the AI-based approach achieved high accuracy similar to manual methods but was significantly faster, suggesting it could improve efficiency in dental practices.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!