Background: Digital pathology operations that precede viewing by a pathologist have a substantial impact on costs and fidelity of the digital image. Scan time and file size determine throughput and storage costs, whereas tissue omission during digital capture ("dropouts") compromises downstream interpretation. We compared how these variables differ across scanners.

Methods: A 212 slide set randomly selected from a gynecologic-gestational pathology practice was used to benchmark scan time, file size, and image completeness. Workflows included the Hamamatsu S210 scanner (operated under default and optimized profiles) and the Leica GT450. Digital tissue dropouts were detected by the aligned overlay of macroscopic glass slide camera images (reference) with images created by the slide scanners whole slide images.

Results: File size and scan time were highly correlated within each platform. Differences in GT450, default S210, and optimized S210 performance were seen in average file size (1.4 vs. 2.5 vs. 3.4 GB) and scan time (93 vs. 376 vs. 721 s). Dropouts were seen in 29.5% (186/631) of successful scans overall: from a low of 13.7% (29/212) for the optimized S210 profile, followed by 34.6% (73/211) for the GT450 and 40.4% (84/208) for the default profile S210 profile. Small dislodged fragments, "shards," were dropped in 22.2% (140/631) of slides, followed by tissue marginalized at the glass slide edges, 6.2% (39/631). "Unique dropouts," those for which no equivalent appeared elsewhere in the scan, occurred in only three slides. Of these, 67% (2/3) were "floaters" or contaminants from other cases.

Conclusions: Scanning speed and resultant file size vary greatly by scanner type, scanner operation settings, and clinical specimen mix (tissue type, tissue area). Digital image fidelity as measured by tissue dropout frequency and dropout type also varies according to the tissue type and scanner. Dropped tissues very rarely (1/631) represent actual specimen tissues that are not represented elsewhere in the scan, so in most cases cannot alter the diagnosis. Digital pathology platforms vary in their output efficiency and image fidelity to the glass original and should be matched to the intended application.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8794031PMC
http://dx.doi.org/10.4103/jpi.jpi_5_21DOI Listing

Publication Analysis

Top Keywords

file size
20
scan time
16
digital pathology
12
tissue
8
tissue dropouts
8
digital image
8
time file
8
glass slide
8
size scan
8
optimized s210
8

Similar Publications

Protocol for quantifying muscle fiber size, number, and central nucleation of mouse skeletal muscle cross-sections using Myotally software.

STAR Protoc

January 2025

Department of Neurology and Neurological Sciences, Stanford University School of Medicine, Stanford, CA 94305, USA; Neurology Service, Veterans Affairs Palo Alto Health Care System, Palo Alto, CA 94304, USA. Electronic address:

Here, we present a protocol for using Myotally, a user-friendly software for fast, automated quantification of muscle fiber size, number, and central nucleation from immunofluorescent stains of mouse skeletal muscle cross-sections. We describe steps for installing the software, preparing compatible images, finding the file path, and selecting key parameters like image quality and size limits. We also detail optional features, such as measuring mean fluorescence.

View Article and Find Full Text PDF

Detecting bolt defects on transmission lines is crucial for ensuring the safe operation of the electrical power system. However, existing methods for detecting bolt defects on transmission lines require higher detection accuracy and smaller model sizes. To address these challenges, this paper proposes a real-time bolt defect detection model based on YOLOv7, named YOLOv7-CWFD.

View Article and Find Full Text PDF

Background: Value-based care payment and delivery models such as the recently implemented Merit-based Incentive Payment System (MIPS) aim to both provide better care for patients and reduce costs of care. Gender disparities across orthopaedic surgery, encompassing reimbursement, industry payments, referrals, and patient perception, have been thoroughly studied over the years, with numerous disparities identified. However, differences in MIPS performance based on orthopaedic surgeon gender have not been comprehensively evaluated.

View Article and Find Full Text PDF

Objective: The present study aimed to evaluate the phase transformation behavior and elemental analysis of thermomechanical-treated nickel-titanium (NiTi) rotary instruments, TruNatomy (Dentsply Sirona), HyFlex CM (coltene, Whaledent), and Neoendo Flex (Orikam healthcare India), using differential scanning calorimetry (DSC), X-ray diffraction (XRD), and energy dispersive X-ray spectrometry.

Materials And Methods: A total of 18 NiTi rotary instruments, TruNatomy, Hyflex CM, Neoendo Flex, taper. 04, size 25 (except TruNatomy, size 26) were selected and were divided into three groups ( = 6).

View Article and Find Full Text PDF

Aim: This study aims to compare and evaluate the fracture resistance of endodontically treated maxillary anterior teeth restored using new postplacement techniques.

Methodology: Forty-five single-rooted maxillary anterior teeth ( = 15) were selected and decoronated at the cementoenamel junction. The working length was determined using a 15K file.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!