Purpose: Accurate assessment of thoracic aortic aneurysm (TAA) growth is important for appropriate clinical management. Maximal aortic diameter is the primary metric that is used to assess growth, but it suffers from substantial measurement variability. A recently proposed technique, termed vascular deformation mapping (VDM), is able to quantify three-dimensional aortic growth using clinical computed tomography angiography (CTA) data using an approach based on deformable image registration (DIR). However, the accuracy and robustness of VDM remains undefined given the lack of ground truth from clinical CTA data, and, furthermore, the performance of VDM relative to standard manual diameter measurements is unknown.

Methods: To evaluate the performance of the VDM pipeline for quantifying aortic growth, we developed a novel and systematic evaluation process to generate 76 unique synthetic CTA growth phantoms (based on 10 unique cases) with variable degrees and locations of aortic wall deformation. Aortic deformation was quantified using two metrics: area ratio (AR), defined as the ratio of surface area in triangular mesh elements and the magnitude of deformation in the normal direction (DiN) relative to the aortic surface. Using these phantoms, we further investigated the effects on VDM's measurement accuracy resulting from factors that influence the quality of clinical CTA data such as respiratory translations, slice thickness, and image noise. Lastly, we compare the measurement error of VDM TAA growth assessments against two expert raters performing standard diameter measurements of synthetic phantom images.

Results: Across our population of 76 synthetic growth phantoms, the median absolute error was 0.063 (IQR: 0.073-0.054) for AR and 0.181 mm (interquartile range [IQR]: 0.214-0.143 mm) for DiN. Median relative error was 1.4% for AR and for DiN at the highest tested noise level (contrast-to-noise ratio [CNR] = 2.66). Error in VDM output increased with slice thickness, with the highest median relative error of 1.5% for AR and 4.1% for DiN at a slice thickness of 2.0 mm. Respiratory motion of the aorta resulted in maximal absolute error of 3% AR and 0.6 mm in DiN, but bulk translations in aortic position had a very small effect on measured AR and DiN values (relative errors ). VDM-derived measurements of magnitude and location of maximal diameter change demonstrated significantly high accuracy and lower variability compared to two expert manual raters ( across all comparisons).

Conclusions: VDM yields an accurate, three-dimensional assessment of aortic growth in TAA patients and is robust to factors such as image noise, respiration-induced translations, and differences in patient position. Further, VDM significantly outperformed two expert manual raters in assessing the magnitude and location of aortic growth despite optimized experimental measurement conditions. These results support validation of the VDM technique for accurate quantification of aortic growth in patients and highlight several important advantages over diameter measurements.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9305918PMC
http://dx.doi.org/10.1002/mp.15496DOI Listing

Publication Analysis

Top Keywords

aortic growth
20
cta data
12
slice thickness
12
growth
11
aortic
11
vdm
9
deformable image
8
image registration
8
taa growth
8
clinical cta
8

Similar Publications

In this study, we investigated gene expression in vitro of human primary Aortic smooth muscle cells (AoSMCs) in response to 9% physiological dynamic stretch over a 4 to 72-h timeframe using RT-qPCR. AoSMC were derived from primary culture and were exposed to continuous cycles of stretch and relaxation at 1 Hz by a computer-controlled Flex Jr.™ Tension System.

View Article and Find Full Text PDF

Abdominal aortic aneurysm (AAA) is a significant vascular condition characterized by the dilation of the abdominal aorta, presenting a substantial risk of rupture and associated high mortality rates. Current management strategies primarily rely on aneurysm diameter and growth rates to predict rupture risk and determine the timing of surgical intervention. However, this approach has limitations, as ruptures can occur in smaller AAAs below surgical thresholds, and many large AAAs remain stable without intervention.

View Article and Find Full Text PDF

Loeys-Dietz syndrome (LDS) is a genetic connective tissue disorder associated with vascular involvement and craniofacial, skeletal, and cutaneous abnormalities. Herein, we describe the case of a 28-year-old female who presented with a pulsatile mass in her abdomen. Imaging studies revealed multiple aneurysms, including a 53-mm abdominal aortic aneurysm (AAA) and tortuosity of the intracranial arterial vasculature.

View Article and Find Full Text PDF

Region-specific biomechanical characterization of ascending thoracic aortic aneurysm of hypertensive patients with bicuspid aortic valves.

Biomech Model Mechanobiol

December 2024

Institute for Biomedical Engineering and Nano Science, Shanghai East Hospital, Tongji University School of Medicine, 500 Zhennan Road, Shanghai, 200331, People's Republic of China.

Hypertension and bicuspid aortic valve (BAV) are key clinical factors that may affect local biomechanical properties of ascending thoracic aortic aneurysms (ATAAs). This study sought to investigate regional differences in biaxial mechanical properties of the ATAAs for the hypertensive patients with BAV. Fresh ATAA samples were harvested from 16 hypertensive patients (age, 66 ± 9 years) undergoing elective aortic surgery.

View Article and Find Full Text PDF

Functional anti-inflammatory mesoporous silica nanoplatform for Synergistic and Targeted abdominal aortic aneurysm treatment.

J Colloid Interface Sci

December 2024

Department of Vascular Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200011, China. Electronic address:

Abdominal aortic aneurysm (AAA) is a chronic inflammation-driven disease characterized by aortic wall destruction and expansion, leading to high morbidity and mortality. However, previous drug treatments for its common risk factors have not achieved favorable results, and the early prevention and treatment is still the main clinical dilemma. Anti-inflammation therapy is a promising therapeutical method targeting its pathogenesis mechanism, but it has not been explored in depth.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!