Objective: Although several commercially available sutureless anastomotic techniques are available, they are not routinely used in neurosurgery. We performed an in vivo flow analysis of end-to-end anastomosis using a microvascular coupler device in rats. We report our first clinical use of the microvascular anastomotic coupler.
Methods: Bilateral rat common carotid arteries (CCAs) were exposed, and a microvascular coupler was used to perform 8 anastomoses. A microflow probe provided quantitative measurement of blood-flow volume. Flow augmentation was assessed with end-to-side anastomoses connecting the distal CCA to the jugular vein (JV). A patient with chronic dominant hemisphere atherosclerotic ischemic disease and progressive symptoms refractory to medical management underwent end-to-end cerebral artery bypass using the microvascular coupler.
Results: Mean preanastomosis flow in the rat CCA was 3.95 ± 0.45 mL/min; this flow was maintained at 3.99 ± 0.24 mL/min on final measurements 54-96 minutes postanastomosis. Total occlusion time for each rat CCA was 12-19 minutes. After end-to-side anastomosis, with proximal and distal JV patent, CCA flow increased 477% to 22.8 ± 3.70 mL/min (P = 0.04, proximal; P = 0.01, distal). After in vivo testing, we successfully used the coupler clinically in a superficial temporal artery-to-middle cerebral artery bypass for dominant hemisphere flow augmentation.
Conclusions: In vivo quantitative flow analysis demonstrated no flow difference between an unaltered artery and artery with end-to-end anastomosis using a microvascular coupler in rats. A 1-mm coupled anastomosis achieved a 4-fold flow increase with low-resistance venous outflow in rats, simulating increased arterial demand. The coupler was successfully used for extracranial-to-intracranial bypass in a patient.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.wneu.2022.01.082 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!