Objectives: Bedside radiographs are usually obtained gridless, without a physical scatter correction grid because of several limitations. Therefore, multiple manufacturers of mobile radiography systems provide the possibility to apply scatter correction software (SC SW) on those images. The purpose of this study was to characterize different series of radiographs-gridless, SC SW, and physical grid-with an image quality assessment algorithm (IQAA). Furthermore, we investigated the potential dose reduction and the correlation between the output of the IQAA and the human observers.

Materials And Methods: We obtained different series of radiographs with an anthropomorphic phantom (multipurpose chest phantom N1 "Lungman," Kyoto Kagaku, Kyoto, Japan). All radiographs were obtained with flat-panel detectors of 5 different manufacturers in a wall bucky system. An IQAA to analyze the radiographs was implemented in our department but was originally developed by the research group of the Duke University Medical Center. Seven physical quantities were calculated by the IQAA: rib-lung contrast (RLcontrast), subdiaphragm-lung contrast (SLcontrast), lung detail (Ldetail), mediastinum detail (Mdetail), lung noise (Lnoise), mediastinum noise (Mnoise), and rib-lung sharpness (RLsharpness). In a proof of concept, the results of the IQAA were validated by 3 experienced radiologists.

Results: Regression coefficients (b) of the linear regression model indicate that the human observer results correlate well with the IQAA (b ≥ 0.89, R2 ≥ 0.83). All manufacturers have SC SW that increases the 7 physical quantities of the gridless images. However, several manufacturers have SC SW that increases the physical metrics to the same level as the physical grid images. The SC SW radiographs obtained with a reduced tube load have an increased level of contrast, detail, sharpness, and noise compared with the gridless images obtained with the higher tube load.

Conclusions: We have proven in a proof of concept that the originally developed IQAA can be used to characterize different series of images of different manufacturers. Based on the physical quantities, SC SW increases the contrast, detail, sharpness, and noise. The experimental results in this study assume a patient dose reduction could be possible when SC SW is applied.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RLI.0000000000000856DOI Listing

Publication Analysis

Top Keywords

scatter correction
12
physical quantities
12
correction software
8
mobile radiography
8
anthropomorphic phantom
8
characterize series
8
dose reduction
8
originally developed
8
proof concept
8
manufacturers increases
8

Similar Publications

The mechanical properties of metal-organic frameworks (MOFs) are of high fundamental and practical relevance. A particularly intriguing technique for determining anisotropic elastic tensors is Brillouin scattering, which so far has rarely been used for highly complex materials like MOFs. In the present contribution, we apply this technique to study a newly synthesized MOF-type material, referred to as GUT2.

View Article and Find Full Text PDF

A multiplexing method based on multidimensional readout method.

Phys Med Biol

January 2025

Institute of Biomedical Engineering, Shenzhen Bay Laboratory, Guangqiao Load, Shenzhen, 518132, CHINA.

To develop and validate a novel multidimensional readout method that significantly reduces the number of readout channels in PET detectors while maintaining high spatial and energy performance. Approach: We arranged a 3×3×4 SiPM array in multiple dimensions and employed row/column/layer summation with a resistor-based splitting circuit. We then applied denoising methods to enhance the peak-to-valley ratio in the decoding map, ensuring accurate crystal-position determination.

View Article and Find Full Text PDF

Ultraportable (UP) X-ray devices are ideal to use in community-based settings, particularly for chest X-ray (CXR) screening of tuberculosis (TB). Unfortunately, there is insufficient guidance on the radiation safety of these devices. This study aims to determine the radiation dose by UP X-ray devices to both the public and radiographers compared to international dose limits.

View Article and Find Full Text PDF

This research examined the distinction between organic and conventional mango fruits, chips, and juice using portable near-infrared (NIR) spectroscopy. A comprehensive analysis was conducted on a sample of 100 mangoes (comprising 50 organic and 50 conventional) utilising a portable NIR spectrometer that spans a wavelength range from 900 to 1700 nm. The mangoes were assessed in their entirety and their juice and chip forms.

View Article and Find Full Text PDF

Machine learning driven benchtop Vis/NIR spectroscopy for online detection of hybrid citrus quality.

Food Res Int

February 2025

State Key Laboratory of Food Science and Resources, Jiangnan University, Wuxi, Jiangsu Province, China; School of Food Science and Technology, Jiangnan University University, Wuxi, Jiangsu Province, China; Collaborative Innovation Center of Food Safety and Quality Control in Jiangsu Province, Jiangnan University, China. Electronic address:

The aim of this study was to explore application of visible and near-infrared (Vis/NIR) spectroscopy combined with machine learning models for SSC and TA prediction of hybrid citrus. The Vis/NIR spectra of samples including navel-region, equator-region and multi-region combination spectra in navel-region and equator-region were collected using a benchtop equipment. The performance of SSC and TA prediction models with different region spectra, including partial least squares (PLS), random forest (RF), k-nearest neighbors (KNN), support vector machine (SVM) and multilayer feedforward neural network (MFNN), was assessed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!