A dose-neutral image quality comparison of different CBCT and CT systems using paranasal sinus imaging protocols and phantoms.

Eur Arch Otorhinolaryngol

Department of Clinical Radiology, Kuopio University Hospital, Puijonlaaksontie 2, POB 100, 70029, Kuopio, Finland.

Published: September 2022

AI Article Synopsis

Article Abstract

Purpose: To compare the image quality produced by equivalent low-dose and default sinus imaging protocols of a conventional dental cone-beam computed tomography (CBCT) scanner, an extremity CBCT scanner and a clinical multidetector computed tomography (MDCT) scanner.

Methods: Three different phantoms were scanned using dose-neutral ultra-low-dose and low-dose sinus imaging protocols, as well as default sinus protocols of each device. Quantified parameters of image quality included modulation transfer function (MTF) to characterize the spatial response of the imaging system, contrast-to-noise ratio, low contrast visibility, image uniformity and Hounsfield unit accuracy. MTF was calculated using the line spread and edge spread functions (LSF and ESF).

Results: The dental CBCT had superior performance over the extremity CBCT in each studied parameter at similar dose levels. The MDCT had better contrast-to-noise ratio, low contrast visibility and image uniformity than the CBCT scanners. However, the CBCT scanners had better resolution compared to the MDCT. Accuracy of HU values for different materials was on the same level between the dental CBCT and MDCT, but substantially poorer performance was observed with the extremity CBCT.

Conclusions: The studied dental CBCT scanner showed superior performance over the studied extremity CBCT scanner when using dose-neutral imaging protocols. In case a dental CBCT is not available, the given extremity CBCT is still a viable option as it provides the benefit of high resolution over a conventional MDCT.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9363284PMC
http://dx.doi.org/10.1007/s00405-022-07271-4DOI Listing

Publication Analysis

Top Keywords

imaging protocols
16
cbct scanner
16
extremity cbct
16
dental cbct
16
image quality
12
cbct
12
sinus imaging
12
default sinus
8
computed tomography
8
contrast-to-noise ratio
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!