Background: Screws and plate are commonly utilized for the fixation of split-type humeral greater tuberosity (GT) fractures. However, the mechanical properties of these 2 types of fixation methods have not been compared directly. The aim of the present study was to evaluate the classic 2 screws fixation with hook locking plate from a mechanical perspective.
Methods: Sixteen synthetic humerii (Sawbones Pacific Research Laboratories, Vashon, WA, USA) were divided into 2 groups. An osteotomy was performed to simulate a split-type GT fracture. Group A (n = 8) was fixed with 2 standard parallel screws. Group B (n = 8) was fixed with a hook plate. Each specimen was tested with traction force at 45° shoulder abduction. Following the 20-N preload, a 500-cycle loading test was applied with a force ranging from 20 to 200 N (valley/peak), and the interfragmental displacement was measured periodically at intervals of 100 cycles. Finally, all the specimens were pulled with destructive force at a rate of 5 mm/min until catastrophic failure.
Results: The hook plate exhibited greater construct stiffness than the screw fixation (63.2 ± 6.1 N/mm vs. 40.9 ± 3.4 N/mm, P < .001). All of the specimens completed the entire cyclic loading test without catastrophic failure, and the fragment displacement after 500 cycles was 0.4 ± 0.2 mm for the hook plate and 2.1 ± 0.3 mm for screw fixation, which was statistically lower in the plate group (P < .001). In terms of failure load, the hook plate group exhibited a significantly greater value than the screw group (770.6 ± 94.6 vs. 427.5 ± 45.1 N/mm, P < .001). The failure modes of both fixation methods were distinct.
Conclusion: In humeral GT fracture fixation, hook plate fixation appears to offer greater construct stiffness and failure load while maintaining fragment stability to resist a physiological traction force. The current study provides support from a mechanical perspective for the clinical application of the hook plate.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jse.2021.12.023 | DOI Listing |
Adv Orthop
December 2024
Department of Orthopedic Surgery, Faculty of Medicine, Al-Azhar University, Cairo, Egypt.
The purpose of this study is to investigate the outcomes of the use of a 2.7 mm semitubular hook plate for internal fixation of unstable metaphyseal ulnar fractures. Between January 2015 and July 2019, 30 consecutive patients with a recent unstable distal ulnar fracture were included in this prospective case series.
View Article and Find Full Text PDFR Soc Open Sci
December 2024
Palaeobiology Research Group, School of Earth Sciences, University of Bristol, Life Sciences Building, Tyndall Avenue, Bristol BS8 1TQ, UK.
Teeth are a key vertebrate innovation; their evolution is generally associated with the origin of jawed vertebrates. However, tooth-like structures already occur in jawless stem-gnathostomes; heterostracans bear denticles and morphologically distinct tubercles on their oral plates. We analysed the histology of the heterostracan denticles and plates to elucidate their morphogenesis and test their homology to the gnathostome oral skeleton.
View Article and Find Full Text PDFCureus
November 2024
Orthopaedic Surgery, Sheikh Zayed Medical College and Hospital, Rahim Yar Khan, PAK.
Background: Dislocation of the acromioclavicular joint (ACJ), accounting for 9%-12% of all shoulder injuries, is a frequent shoulder problem. Clavicular hook plates have proven to be a successful implant choice for surgical management. The benefit of this method is that it preserves the diarthrosis of the ACJ while reducing dislocation.
View Article and Find Full Text PDFJ Foot Ankle Surg
November 2024
Beijing Chaoyang Hospital of Capital Medical University, Beijing 100020, China. Electronic address:
J Shoulder Elbow Surg
November 2024
Department of Orthopedic Surgery, Hangzhou Xiaoshan Traditional Chinese Medical Orthopedic Hospital, Hangzhou, China.
Background: Subacromial osteolysis is a typical complication following hook plate fixation for acromioclavicular (AC) dislocation. Many factors can affect the occurrence and progression of subacromial osteolysis (SAO). The objective of this study was to investigate the predictive value of the implant removal time and acromion-hook angle for subacromial osteolysis following hook plate fixation for AC dislocation.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!