Department of Orthopaedic Surgery, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan; Skeleton Materials and Bio-compatibility Core Lab, Research Center of Clinical Medicine, National Cheng Kung University Hospital, College of Medicine, National Cheng Kung University, Tainan, Taiwan; Medical Device Innovation Center, National Cheng Kung University, Tainan, Taiwan. Electronic address:

Published: June 2022

Background: Screws and plate are commonly utilized for the fixation of split-type humeral greater tuberosity (GT) fractures. However, the mechanical properties of these 2 types of fixation methods have not been compared directly. The aim of the present study was to evaluate the classic 2 screws fixation with hook locking plate from a mechanical perspective.

Methods: Sixteen synthetic humerii (Sawbones Pacific Research Laboratories, Vashon, WA, USA) were divided into 2 groups. An osteotomy was performed to simulate a split-type GT fracture. Group A (n = 8) was fixed with 2 standard parallel screws. Group B (n = 8) was fixed with a hook plate. Each specimen was tested with traction force at 45° shoulder abduction. Following the 20-N preload, a 500-cycle loading test was applied with a force ranging from 20 to 200 N (valley/peak), and the interfragmental displacement was measured periodically at intervals of 100 cycles. Finally, all the specimens were pulled with destructive force at a rate of 5 mm/min until catastrophic failure.

Results: The hook plate exhibited greater construct stiffness than the screw fixation (63.2 ± 6.1 N/mm vs. 40.9 ± 3.4 N/mm, P < .001). All of the specimens completed the entire cyclic loading test without catastrophic failure, and the fragment displacement after 500 cycles was 0.4 ± 0.2 mm for the hook plate and 2.1 ± 0.3 mm for screw fixation, which was statistically lower in the plate group (P < .001). In terms of failure load, the hook plate group exhibited a significantly greater value than the screw group (770.6 ± 94.6 vs. 427.5 ± 45.1 N/mm, P < .001). The failure modes of both fixation methods were distinct.

Conclusion: In humeral GT fracture fixation, hook plate fixation appears to offer greater construct stiffness and failure load while maintaining fragment stability to resist a physiological traction force. The current study provides support from a mechanical perspective for the clinical application of the hook plate.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jse.2021.12.023DOI Listing

Publication Analysis

Top Keywords

hook plate
12
greater tuberosity
8
tuberosity fractures
8
group n = 8
8
n = 8 fixed
8
plate
5
biomechanical comparisons
4
hook
4
comparisons hook
4
plate screw
4

Similar Publications

The purpose of this study is to investigate the outcomes of the use of a 2.7 mm semitubular hook plate for internal fixation of unstable metaphyseal ulnar fractures. Between January 2015 and July 2019, 30 consecutive patients with a recent unstable distal ulnar fracture were included in this prospective case series.

View Article and Find Full Text PDF

Morphogenesis of pteraspid heterostracan oral plates and the evolutionary origin of teeth.

R Soc Open Sci

December 2024

Palaeobiology Research Group, School of Earth Sciences, University of Bristol, Life Sciences Building, Tyndall Avenue, Bristol BS8 1TQ, UK.

Teeth are a key vertebrate innovation; their evolution is generally associated with the origin of jawed vertebrates. However, tooth-like structures already occur in jawless stem-gnathostomes; heterostracans bear denticles and morphologically distinct tubercles on their oral plates. We analysed the histology of the heterostracan denticles and plates to elucidate their morphogenesis and test their homology to the gnathostome oral skeleton.

View Article and Find Full Text PDF

Background: Dislocation of the acromioclavicular joint (ACJ), accounting for 9%-12% of all shoulder injuries, is a frequent shoulder problem. Clavicular hook plates have proven to be a successful implant choice for surgical management. The benefit of this method is that it preserves the diarthrosis of the ACJ while reducing dislocation.

View Article and Find Full Text PDF
Article Synopsis
  • The study examined the best treatment options for displaced or comminuted fractures of the fifth metatarsal base by comparing three methods: ulnar hook plate fixation, immobilization, and early functional treatment.
  • Results showed that while the plate fixation group had better AOFAS and FADI scores at 1, 3, and 6 months, all treatments had similar outcomes after 12 months, with early functional treatment leading to quicker recovery times.
  • The study concluded that early functional treatment is recommended due to its effectiveness and rapid return to daily activities, despite some complications in the plate fixation group.
View Article and Find Full Text PDF

Background: Subacromial osteolysis is a typical complication following hook plate fixation for acromioclavicular (AC) dislocation. Many factors can affect the occurrence and progression of subacromial osteolysis (SAO). The objective of this study was to investigate the predictive value of the implant removal time and acromion-hook angle for subacromial osteolysis following hook plate fixation for AC dislocation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!