A PHP Error was encountered

Severity: Warning

Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests

Filename: helpers/my_audit_helper.php

Line Number: 176

Backtrace:

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML

File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016

File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global

File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword

File: /var/www/html/index.php
Line: 316
Function: require_once

Association between Previous Pelvic Radiation and All-Cause and Cause-Specific Failure of Replacement Artificial Urinary Sphincters. | LitMetric

AI Article Synopsis

  • The study investigated how previous radiation therapy impacts the lifespan and failure rates of artificial urinary sphincters (AUSs) in men.
  • It found that men with a history of radiation experienced significantly shorter times to device failure compared to those without (1.4 years vs. 3.5 years).
  • Specifically, prior radiation dramatically increased the risk of erosion or infection related to the AUS but did not affect the risk of urethral atrophy or mechanical failures.

Article Abstract

Purpose: In order to accurately characterize how a history of radiation therapy affects the lifespan of replacement artificial urinary sphincters (AUSs), all possible sources of device failure must be considered. We assessed the competing risks of device failure based on radiation history in men with replacement AUSs.

Materials And Methods: We identified men who had a replacement AUS in a single institutional, retrospective database. To assess survival from all-cause device failure based on radiation history and other factors, we conducted Kaplan-Meier, Cox proportional-hazards and competing risks analyses.

Results: Among 247 men who had a first replacement AUS, men with a history of radiation had shorter time to all-cause device failure (median 1.4 vs 3.5 years for men with radiation vs without radiation history, p=0.02). On multivariable Cox-proportional hazards analysis, previous radiation was associated with increased risk of all-cause device failure (HR: 2.12, 95% CI: 1.30-3.43, p=0.002) On multivariable cause-specific hazards analysis, prior radiation was associated with a higher risk of erosion/infection (HR: 7.57, 95% CI: 2.27-25.2, p <0.001), but was not associated with risk of urethral atrophy (p=0.5) or mechanical failure (p=0.15).

Conclusions: Among men with a replacement AUS, a history of pelvic radiation was associated with shorter time to device failure of any cause. Radiation was also specifically associated with a sevenfold increase in the risk of erosion or infection of replacement AUS, but not with urethral atrophy or mechanical failure. Patients with a replacement AUS should be appropriately counseled on how radiation history may impact outcomes of future revisions.

Download full-text PDF

Source
http://dx.doi.org/10.1097/JU.0000000000002433DOI Listing

Publication Analysis

Top Keywords

device failure
20
radiation history
12
men replacement
12
all-cause device
12
radiation
9
replacement artificial
8
artificial urinary
8
urinary sphincters
8
history radiation
8
competing risks
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!