Introduction: In this feasibility study, we hypothesize that the evaluation of cervical biomechanical strength can be improved if cervical length measurement is supplemented with quantitative elastography, which is a technique based on conventional ultrasound elastography combined with a force-measuring device. Our aims were to: (a) develop a force-measuring device; (b) introduce a cervical elastography index (CEI) and a cervical strength index (CSI; defined as cervical length × CEI); (c) evaluate how these indexes assess the cervical softening that takes place during normal pregnancy; and (d) how these indexes predict the cervical dilatation time from 4 to 10 cm.
Material And Methods: An electronic force-measuring device was mounted on the handle of the transvaginal probe, allowing for force measurement when conducting elastography. The study group concerned with normal cervical softening included 44 unselected pregnant women. Outcomes were CEI and CSI at different gestational ages. The study group for labor induction included 26 singleton term pregnant women admitted for labor induction. Outcome was defined as cervical dilatation time from 4 to 10 cm. Elastography measured the changes in mean gray value (intensity) during manual compressions. Region of interest was set within the anterior cervical lip.
Results: We found that the mean of all variables regarding cervical softening decreased from early to late pregnancy: ie cervical length from 34 to 29 mm, CEI from 0.17 to 0.11 N, and CSI from 5.9 to 3.1 N mm. Moreover, the cervical dilatation time during labor induction was associated with CEI, although not statistically significantly (area under the ROC curve of 0.67), but not with the Bishop score, the cervical length, or the CSI.
Conclusions: We propose that quantitative elastography based on changes in the intensity of the B-mode ultrasound recording, in combination with a force-measuring device on the handle of the vaginal probe, deserves further investigation as an approach for evaluation of cervical biomechanical strength.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC9564675 | PMC |
http://dx.doi.org/10.1111/aogs.14309 | DOI Listing |
Objective: Limitations in human kinematics during cochlear implantation induce pressure transients and increased forces on intracochlear structures. Herein, we present a novel head-mounted surgical tool designed for the motorized insertion of cochlear implant electrode arrays. The tool integrates a force measurement feature to overcome the lack of haptic feedback in current robotic solutions.
View Article and Find Full Text PDFCureus
May 2024
Oral Medicine and Radiology, Dental Institute, Rajendra Institute of Medical Sciences, Ranchi, IND.
Introduction: It is critical to measure the maximum voluntary bite force of patients receiving restorative dentistry. A new device known as "BYTE" has been developed indigenously to measure bite force in humans. The purpose of this study is to evaluate the BYTE device's consistency and accuracy in a lab setting.
View Article and Find Full Text PDFMedicina (Kaunas)
March 2024
Faculty of Sports Sciences, Yozgat Bozok University, Yozgat 66200, Turkey.
: The accurate prediction of strength development relies on identifying the most appropriate measurement methods. This study compared diverse strength measurement techniques to assess their effectiveness in predicting strength development. Participants were taekwondo athletes competing at the red-black belt level or above.
View Article and Find Full Text PDFSensors (Basel)
February 2024
School of Electronic & Information Engineering, Suzhou University of Science and Technology, Suzhou 215009, China.
The choice of torque curve in lower-limb enhanced exoskeleton robots is a key problem in the control of lower-limb exoskeleton robots. As a human-machine coupled system, mapping from sensor data to joint torque is complex and non-linear, making it difficult to accurately model using mathematical tools. In this research study, the knee torque data of an exoskeleton robot climbing up stairs were obtained using an optical motion-capture system and three-dimensional force-measuring tables, and the inertial measurement unit (IMU) data of the lower limbs of the exoskeleton robot were simultaneously collected.
View Article and Find Full Text PDFSensors (Basel)
February 2024
Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China.
To guarantee the accuracy of sophisticated equipment in spacecraft, it is essential to evaluate the dynamic forces of vibration sources. In contrast to conventional rigid-based measuring approaches, a method for predicting the interference of dynamic forces from large sources on spacecraft considering vibration coupling is proposed. In addition, a flexible-based dynamic force measuring platform capable of withstanding large masses and mounting large-volume vibration sources is designed.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!