Background: Since the beginning of the pandemic, clinicians and researchers have been searching for alternative tests to improve the screening and diagnosis of the SARS-CoV-2 infection. Currently, the gold standard for virus identification is the nasopharyngeal (NP) swab. Saliva samples, however, offer clear, practical, and logistical advantages but due to a lack of collection, transport, and storage solutions, high-throughput saliva-based laboratory tests are difficult to scale up as a screening or diagnostic tool. With this study, we aimed to validate an intralaboratory molecular detection method for SARS-CoV-2 on saliva samples collected in a new storage saline solution, comparing the results to NP swabs to determine the difference in sensitivity between the two tests.

Methods: In this study, 156 patients (cases) and 1005 asymptomatic subjects (controls) were enrolled and tested simultaneously for the detection of the SARS-CoV-2 viral genome by RT-PCR on both NP swab and saliva samples. Saliva samples were collected in a preservative and inhibiting saline solution (Biofarma Srl). Internal method validation was performed to standardize the entire workflow for saliva samples.

Results: The identification of SARS-CoV-2 conducted on saliva samples showed a clinical sensitivity of 95.1% and specificity of 97.8% compared to NP swabs. The positive predictive value (PPV) was 81% while the negative predictive value (NPV) was 99.5%. Test concordance was 97.6% (Cohen's Kappa = 0.86; 95% CI 0.81-0.91). The LoD of the test was 5 viral copies for both samples.

Conclusions: RT-PCR assays conducted on a stored saliva sample achieved similar performance to those on NP swabs, and this may provide a very effective tool for population screening and diagnosis. Collection of saliva in a stabilizing solution makes the test more convenient and widely available; furthermore, the denaturing properties of the solution reduce the infective risks belonging to sample manipulation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8759915PMC
http://dx.doi.org/10.1155/2022/6478434DOI Listing

Publication Analysis

Top Keywords

saliva samples
20
diagnosis sars-cov-2
8
sars-cov-2 infection
8
screening diagnosis
8
saliva
8
swab saliva
8
samples collected
8
saline solution
8
sars-cov-2
5
samples
5

Similar Publications

This study compared the dynamics of SARS-CoV-2 viral shedding in saliva between wild-type virus-infected and Omicron-infected household cohorts. Pre-existing immunity in participants likely shortens the viral RNA shedding duration and lowers viral load peaks. Frequent saliva sampling can be a convenient tool to study viral load dynamics.

View Article and Find Full Text PDF

Dogs can discriminate between people infected with SARS-CoV-2 from those uninfected, although their results vary depending on the settings in which they are exposed to infected individuals or samples of urine, sweat or saliva. This variability likely depends on the viral load of infected people, which may be closely associated with physiological changes in infected patients. Determining this viral load is challenging, and a practical approach is to use the cycle threshold (Ct) value of a RT-qPCR test.

View Article and Find Full Text PDF

Nowadays, gastroesophageal reflux disease (GERD) has emerged as one of the major hazards to the health of the upper gastrointestinal tract, and there is an urgent need for a low-cost, user-friendly, and non-invasive detection method. Herein, a paper-based sensor (CP sensor) for the non-invasive screening of GERD is proposed. The sensor is structured as a specially shaped cellulose paper strip embedded with fluorescent colloids, which are self-assembled from a cleavable synthetic fluorescent polymer (P4).

View Article and Find Full Text PDF

Background: Oral microbiome homeostasis is important for children's health, and microbial community is affected by anesthetics. The application of anesthetics in children's oral therapy has become a relatively mature method. This study aims to investigate the effect of different anesthesia techniques on children's oral microbiota.

View Article and Find Full Text PDF

Background: Examples of group B red cells that react weakly or not at all with anti-B have been described. Subgroups of B such as B, B, B, and B are rare and are less frequently reported. We studied the frequency of subgroups of B in our healthy blood donor population and serologically characterized and differentiated these subgroups.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!