Objectives: Increase the yield and purity of cell-free DNA (cfDNA) extracted from plasma for non-invasive prenatal testing (NIPT) as inefficiencies in this extraction and purification can dramatically affect the sensitivity and specificity of the test.

Methods: This work integrates cfDNA extraction from plasma with a microfluidic chip platform by combining magnetic bead-based extraction and electroosmotic flow on the microfluidic chip. Various wash buffers and voltage conditions were simulated using COMSOL Multiphysics Modeling and tested experimentally.

Results: When performing the first wash step of this assay on the microfluidic chip with 300 V applied across the channel there was a six-fold increase in the A /A ratio showing a significant improvement (p value 0.0005) in the purity of the extracted sample all while maintaining a yield of 68.19%. These values are critical as a high yield results in more sample to analyze and an increase in A /A ratio corresponds to a decrease in salt contaminants such as guanidinium thiocyanate which can interfere with downstream processes during DNA library preparation and potentially hinder the NIPT screening results.

Conclusions: This technique has the potential to improve NIPT outcomes and other clinically relevant workflows that use cfDNA as an analyte such as cancer detection.

Download full-text PDF

Source
http://dx.doi.org/10.1002/pd.6092DOI Listing

Publication Analysis

Top Keywords

microfluidic chip
12
extraction plasma
8
plasma non-invasive
8
non-invasive prenatal
8
prenatal testing
8
increase ratio
8
microfluidic
4
microfluidic platform
4
platform high-purity
4
high-purity cell
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!