Purpose: To develop and assess the accuracy of a mixed reality (MR) needle guidance application on smartglasses.

Materials And Methods: An MR needle guidance application on HoloLens2, without pre-procedural CT image reconstruction or import by manually matching the spatial and MR coordinate systems, was developed. First, the accuracy of the target locations in the image overlay at 63 points arranged on a 45 × 35 × 21 cm box and needle angles from 0° to 80°, placed using the MR application, was verified. The needle placement errors from 12 different entry points in a phantom by seven operators (four physicians and three non-physicians) were compared using a linear mixed model between the MR guidance and conventional methods using protractors.

Results: The average errors of the target locations and needle angles placed using the MR application were 5.9 ± 2.6 mm and 2.3 ± 1.7°, respectively. The average needle insertion error using the MR guidance was slightly smaller compared to that using the conventional method (8.4 ± 4.0 mm vs. 9.6 ± 5.1 mm, p = 0.091), particularly in the out-of-plane approach (9.6 ± 3.5 mm vs. 12.3 ± 4.6 mm, p = 0.003). The procedural time was longer with MR guidance than with the conventional method (412 ± 134 s vs. 219 ± 66 s, p < 0.001).

Conclusion: MR needle guidance without pre-procedural CT image import is feasible when matching coordinate systems, and the accuracy of needle insertion is slightly better than that of the conventional method.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00270-021-03029-3DOI Listing

Publication Analysis

Top Keywords

needle guidance
12
guidance application
12
mixed reality
8
reality needle
8
pre-procedural image
8
import manually
8
manually matching
8
coordinate systems
8
target locations
8
needle angles
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!