Study Design: Experimental study.

Objective: In this study, the ambient temperature of a radiofrequency (RF) electrode tip was compared and analyzed in terms of products, mode, flow quantity, and flow rate.

Summary Of Background Data: Endoscopic spine surgery is a widely used operation for degenerative lumbar stenosis and herniated lumbar disc. To perform endoscopic spine surgery, dedicated instruments like a RF generator and electrode are essential.

Methods: An evaluation system capable of measuring temperature under equal conditions at a certain distance from the electrode tip was manufactured. The distance between the electrode tip and the temperature sensor was set to 1, 5, and 10 mm. The flow quantities of 0, 50, 100, and 150 mL/min and the flow rates of 0, 0.20, 0.53, and 0.80 m/s were compared and statistically analyzed.

Results: The temperatures measured in the experiments conducted on the four combinations of RF device showed similar values, and showed differences according to the characteristics of each mode of the RF. As the distance between the electrode tip and the temperature sensor increased, the temperature decreased, and as flow quantity or flow rate increased, the temperature decreased. The maximum temperatures differed significantly according to flow quantity, between flow quantities of 0 and 100 mL/min (P  = 0.03) and between 0 and 150 mL/min (P ≤ 0.01). The maximum temperatures also differed significantly between the flow rate of 0 m/s, and the flow rates of 0.20, 0.53, and 0.80 m/s, with P ≤ 0.01 in all three comparisons.

Conclusion: This is the first study in which we made a customized RF temperature evaluation system and verified the temperature changes in various environments. When irrigation was performed, we could confirm that the maximum temperature was less than 60°C. Irrigation is considered essential in endoscopic spine surgery.

Level Of Evidence: 3.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BRS.0000000000004320DOI Listing

Publication Analysis

Top Keywords

endoscopic spine
16
spine surgery
12
flow quantity
12
quantity flow
12
distance electrode
12
flow
10
temperature
9
evaluation system
8
electrode temperature
8
temperature sensor
8

Similar Publications

Objective: Percutaneous Endoscopic Transforaminal Discectomy (PETD) is recognized as the leading surgical intervention for lumbar disc herniation (LDH). Moreover, Body Mass Index (BMI) has been established as an independent risk factor for disc reherniation post-PETD. Furthermore, there is a lack of studies investigating the biomechanical changes in the disc post-PETD in relation to diverse BMI levels.

View Article and Find Full Text PDF

Study Design: Retrospective cohort study.

Objective: to analyze trends in PROMs improvement and recovery kinetics following transforaminal endoscopic lumbar discectomy and foraminotomy (TELD).

Summary Of Background Data: As TELDs become an increasingly common alternative to fusions for lateral disc herniations, it is important to understand patients' postoperative recovery timelines to manage patient expectations.

View Article and Find Full Text PDF

Left T7 paravertebral nerve blockade activate the α7nAChR-Dependent CAP in patients undergoing thoracoscopic lobectomy: a prospective controlled study.

BMC Anesthesiol

December 2024

Department of Anesthesiology and Surgery, The First Affiliated Hospital, Jiangxi Medical College, Nanchang University, Nanchang, Jiangxi, 330001, China.

Objective: This study aimed to observe the impact of Tthoracic paravertebral nerve blockade(TPVB) at left T7 level on the α7nAChR-dependent cholinergic anti-inflammatory pathway in patients undergoing thoracoscopic lobectomy.

Methods: Scheduled thoracoscopic lung surgery patients at the First Affiliated Hospital of Nanchang University from August to September 2023 were divided into two groups according to the surgical site. The experimental group underwent left T7 paravertebral nerve blockade (LTPVB group), while the control group underwent right T7 paravertebral nerve blockade (RTPVB group).

View Article and Find Full Text PDF

Objectives: The study aimed to evaluate the hidden blood loss (HBL) and its possible risk factors in patients with lumbar disc herniation undergoing percutaneous endoscopic lumbar discectomy (PELD) via posterolateral approach.

Patients And Methods: The clinical data of 170 lumbar disc herniation patients (101 males, 69 females; mean age: 57.7±18.

View Article and Find Full Text PDF

Objectives: The study aimed to evaluate the hidden blood loss (HBL) and its possible risk factors in patients undergoing percutaneous endoscopic cervical discectomy (PECD) via posterior approach to better guide the management of perioperative anemia in patients.

Patients And Methods: The study retrospectively analyzed the clinical data of 60 patients (33 males, 27 females; mean age: 55.3±7.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!