Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Improving resection accuracy and eliminating outliers in total knee arthroplasty (TKA) is important to improving patient outcomes regardless of alignment philosophy. Robotic-assisted surgical systems improve resection accuracy and reproducibility compared to conventional instrumentation. Some systems require preoperative imaging while others rely on intraoperative anatomic landmarks. We hypothesized that the alignment accuracy of a novel image-free robotic-assisted surgical system would be equivalent or better than conventional instrumentation with fewer outliers.
Methods: Forty cadaveric specimens were used in this study. Five orthopedic surgeons performed 8 bilateral TKAs each, using the VELYS Robotic-Assisted System (DePuy Synthes) and conventional instrumentation on contralateral knees. Pre-resection and postresection computed tomography scans, along with optical scans of the implant positions were performed to quantify resection accuracies relative to the alignment targets recorded intraoperatively.
Results: The robotic-assisted cohort demonstrated smaller resection errors compared to conventional instrumentation in femoral coronal alignment (0.63° ± 0.50° vs 1.39° ± 0.95°, P < .001), femoral sagittal alignment (1.21° ± 0.90° vs 3.27° ± 2.51°, P < .001), and tibial coronal alignment (0.93° ± 0.72° vs 1.65° ± 1.29°, P = .001). All other resection angle accuracies were equivalent. Similar improvements were found in the femoral implant coronal alignment (0.89° ± 0.82° vs 1.42° ± 1.15°, P = .011), femoral implant sagittal alignment (1.51° ± 1.08° vs 2.49° ± 2.10°, P = .006), and tibial implant coronal alignment (1.31° ± 0.84° vs 2.03° ± 1.44°, P = .004). The robotic-assisted cohort had fewer outliers (errors >3°) for all angular resection alignments.
Conclusion: This in vitro study demonstrated that image-free robotic-assisted TKA can improve alignment accuracy compared to conventional instrumentation and reduce the incidence of outliers.