Background: Real-time ultrasound (US) guidance facilitates central venous catheterization in intensive care unit (ICU). New magnetic needle-pilot devices could improve efficiency and safety of central venous catheterization. This simulation trial was aimed at comparing venipuncture with a new needle-pilot device to conventional US technique.

Methods: In a prospective, randomized, simulation trial, 51 ICU physicians and residents cannulated the right axillary vein of a human torso mannequin with standard US guidance and with a needle-pilot system, in a randomized order. The primary outcome was the time from skin puncture to successful venous cannulation. The secondary outcomes were the number of skin punctures, the number of posterior wall puncture of the axillary vein, the number of arterial punctures, the number of needle redirections, the failure rate, and the operator comfort.

Results: Time to successful cannulation was shorter with needle-pilot US-guided technique (22 s (interquartile range (IQR) = 16-42) vs 25 s (IQR = 19-128); median of difference (MOD) = -9 s (95%-confidence interval (CI) -5, -22),  < 0.001). The rates of skin punctures, posterior wall puncture of axillary vein, and needle redirections were also lower ( < 0.01). Comfort was higher in needle-pilot US-guided group on a 11-points numeric scale (8 (IQR = 8-9) vs 6 (IQR = 6-8),  < 0.001).

Conclusions: In a simulation model, US-guided axillary vein catheterization with a needle-pilot device was associated with a shorter time of successful cannulation and a decrease in numbers of skin punctures and complications. The results plea for investigating clinical performance of this new device.

Download full-text PDF

Source
http://dx.doi.org/10.1177/11297298211063705DOI Listing

Publication Analysis

Top Keywords

axillary vein
12
ultrasound guidance
8
prospective randomized
8
needle-pilot device
8
central venous
8
venous catheterization
8
simulation trial
8
punctures number
8
needle-pilot
5
vein catheterization
4

Similar Publications

The plane running between two adjacent pulmonary segments consists of a very thin layer of connective tissue through which the pulmonary vein also runs. To perform an anatomically correct segmentectomy, this segmental plane needs to be divided. Before the operation, the locations of vessels and bronchi are confirmed by three-dimensional computed tomography.

View Article and Find Full Text PDF

Unlabelled: Upper extremity deep vein thrombosis (UEDVT) is relatively rare, and much less as an initial presentation of systemic lupus erythematosus (SLE). Primary UEDVT should be considered in individuals with unilateral arm swelling where the brachial, axillary, and subclavian veins are frequently involved. SLE is a chronic autoimmune disease that predominantly affects women of childbearing age and of African descent.

View Article and Find Full Text PDF

Purpose: To validate the Axillary Reverse Mapping (ARM) technique with indocyanine green (ICG), focusing on the detection rate and the procedure's feasibility. The predictive factors for metastatic involvement of ARM nodes are also analyzed to define the target population for ARM indication.

Methods: This prospective, observational, non-randomized study of patients with breast cancer included patients with an indication for axillary lymph node dissection (ALND) performed between June 2021 and June 2023.

View Article and Find Full Text PDF

Introduction: Ultrasound (US)-guided axillary vein puncture is a safe and effective approach for cardiac implantable electronic device (CIED) implantation, and it is highly recommended by the current consensus document. However, only reports on small populations are available in the current literature regarding the comparison of this technique with other traditional approaches (subclavian vein blind puncture and cephalic vein surgical cutdown).

Purpose: We aimed to assess the effectiveness and safety of US- guided axillary vein puncture using a microintroducer kit for CIED implantation as compared to the aforementioned traditional approaches.

View Article and Find Full Text PDF

Hepatectomy with Hepatic Vein Resection and Reconstruction Under Total Vascular Exclusion and Venous Drainage via a Venovenous Bypass: An Additional Approach for Complex Hepatectomies.

Ann Surg Oncol

December 2024

Hepato-Pancreato-Biliary Surgery and Liver Transplantation, Pôle des Pathologies Hépatiques et Digestives, Hôpital de Hautepierre-Hôpitaux Universitaires de Strasbourg, Université de Strasbourg, Strasbourg, France.

Background: Total vascular exclusion (TVE) with liver hypothermic perfusion under venovenous bypass (VVB) is usually needed to perform hepatectomy with Inferior vena cava and hepatic veins resection-reconstruction. An alternative technique is represented by liver resection under intermittent pedicular clamping, IVC total clamping and VVB, without cold perfusion and liver outflow drainage through the VVB. PATIENTS AND METHODS: The patient is a 60-year-old woman with past medical history of right hepatectomy for leiomyosarcoma 14 years previously.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!