Introduction: The aims of study were to assess: 1) performance specifications of Atellica 1500, 2) comparability of Atellica 1500 and Iris, 3) the accuracy of both analysers in their ability to detect bacteria.

Materials And Methods: Carryover, linearity, precision, reproducibility, and limit of blank (LoB) verification were evaluated for erythrocyte and leukocyte counts. ICSH 2014 protocol was used for estimation of carryover, CLSI EP15-A3 for precision, and CLSI EP17 for LoB verification. Comparison for quantitative parameters was evaluated by Bland-Altman plot and Passing-Bablok regression. Qualitative parameters were evaluated by Weighted kappa analysis. Sixty-five urine samples were randomly selected and sent for urine culture which was used as reference method to determine the accuracy of bacteria detection by analysers.

Results: Analytical specifications of Atellica 1500 were successfully verified. Total of 393 samples were used for qualitative comparison, while 269 for sediment urinalysis. Bland-Altman analysis showed statistically significant proportional bias for erythrocytes and leukocytes. Passing-Bablok analysis for leukocytes pointed to significant constant and minor proportional difference, while it was not performed for erythrocytes due to significant data deviation from linearity. Kappa analysis resulted in the strongest agreements for pH, ketones, glucose concentrations and leukocytes, while the poorest agreement for bacteria. The sensitivity and specificity of bacteria detection were: 91 (59-100)% and 76 (66-87)% for Atellica 1500 and 46 (17-77)% and 96 (87-100)% for Iris.

Conclusion: There are large differences between Atellica 1500 and Iris analysers, due to which they are not comparable and can not be used interchangeably. While there was no difference in specificity of bacteria detection, Iris analyser had greater sensitivity.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8672386PMC
http://dx.doi.org/10.11613/BM.2022.010701DOI Listing

Publication Analysis

Top Keywords

atellica 1500
24
bacteria detection
12
urine culture
8
specifications atellica
8
1500 iris
8
lob verification
8
parameters evaluated
8
kappa analysis
8
specificity bacteria
8
0
6

Similar Publications

Objectives: Autoverification increases the efficiency of laboratories. Laboratories accredited according to ISO 15189:2022 need to validate their processes, including autoverification, and assess the associated risks to patient safety. The aim of this study was to propose a systematic verification algorithm for autoverification and to assess its potential risks.

View Article and Find Full Text PDF
Article Synopsis
  • The Atellica® VTLi high sensitivity cardiac troponin-I assay is being studied to improve diagnosis of myocardial infarction in patients with suspected acute coronary syndrome in the emergency department.
  • The study will include around 1,500 patients who will provide blood samples at various times and undergo specific scoring, with results being kept blind to the clinical team.
  • The researchers aim to demonstrate that the performance of this point-of-care test is comparable to traditional lab-based tests, delivering results in about 8 minutes.
View Article and Find Full Text PDF

Urinary tract infection (UTI) is one of the most common diseases occurring in both hospitalized and community subjects. Urine culture is the gold standard test for the diagnosis of UTI, but approximately 80% are negative. The aim of this study was to evaluate the performance of the automated urinalysis system Atellica® 1500 (Siemens Healthineers, Erlangen, Germany) as screening tool for ruling out UTI.

View Article and Find Full Text PDF

Non-linearity in lipase assays: A multicentric comparison on different analysers.

Clin Biochem

April 2023

Service de Biochimie et Biologie Moléculaire, Laboratoire de Biologie Médicale Multi-Sites (LBMMS), Hospices Civils de Lyon (HCL), F-69000 Lyon, France.

Objectives: Non-linearity in lipase assays and the ensuing gaps in results distribution have been described on Roche analysers, but have yet to be studied on other analysers.

Design And Methods: Eighteen lithium-heparinized plasma pools of lipase activities decreasing from 1700 to <4 U/L were prepared for multicentric evaluation on several analysers. Non-linearity was modelled as the difference between the polynomial regression of lipase activities depending on relative dilutions over the primary measuring range, and the linear regression of the same variables above the manufacturer's limit of linearity (MLL).

View Article and Find Full Text PDF

Introduction: The Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!