Optical Bench Analysis of 2 Depth of Focus Intraocular Lenses.

Biomed Hub

Accuratis, Practice for Refractive Eye Surgery, Ulm, Germany.

Published: September 2021

Background: The aim of the study was to analyze the objective optical properties of 2 enhanced depth of focus (EDoF) intraocular lenses (IOLs) using optical bench analysis.

Methods: This experimental study investigates 2 new EDoF IOLs, the Alcon AcrySof IQ Vivity and the Bausch & Lomb LuxSmart Crystal, on the optical bench, using OptiSpheric IOL PRO2 (Trioptics, Germany) in order to assess the optical quality according to ISO 11979 with ISO-2 Cornea. IOLs (power 22.0 D) were evaluated regarding modulation transfer function (MTF) at 50 lp/mm and Strehl ratio (SR) using a 3.0-mm and a 4.5-mm aperture. In addition, wavefront measurements were obtained using WaveMaster® IOL 2 device (Trioptics, Germany), and USAF targets were analyzed.

Results: Centered: the MTF (mean) at 50 lp/mm (AcrySof IQ Vivity/LuxSmart Crystal) with 3.0 mm aperture was 0.250/0.257 and with 4.5 mm aperture 0.202/0.243. The SR (mean) with 3.0 mm aperture was 0.261/0.355 and with 4.5 mm aperture 0.176/0.206. Decentered by 1 mm: the MTF (mean) at 50 lp/mm (AcrySof IQ Vivity/LuxSmart Crystal) with 3.0 mm aperture was 0.266/0.247 and with 4.5 mm aperture 0.126/0.215. The SR (mean) with 3.0 mm aperture was 0.272/0.234 and with 4.5 mm aperture 0.133/0.183. Tilted by 5 degree: the MTF (mean) at 50 lp/mm (AcrySof IQ Vivity/LuxSmart Crystal) with 3.0 mm aperture was 0.221/0.360 and with 4.5 mm aperture 0.214/0.229. The SR (mean) with 3.0 mm aperture was 0.232/0.428 and with 4.5 mm aperture 0.225/0.229. The simulated visual function using USAF test targets showed corresponding qualitative results. Wavefront measurements proved a complex optical design. Higher order aberrations in the central part of the optics were modulated up to the 10th order to enhance the range of functional vision to near distance, leaving the peripheral parts of the optics aberration free or as aberration correcting.

Conclusion: The diversity of EDOF IOLs, their optics, and their respective impact on the vision quality must be understood in order to select the appropriate IOL in each individual case. This analysis of new, innovative IOL optics based on increased negative spherical aberration may help the ophthalmic surgeon to select the IOL which meets the individual requirements of the patient for best postoperative outcomes. It seems that there is no perfect IOL that is equally suitable for all patients, but the right choice is an individual, customized approach dealing with patients' expectations.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC8613612PMC
http://dx.doi.org/10.1159/000519139DOI Listing

Publication Analysis

Top Keywords

mtf lp/mm
16
aperture
13
optical bench
12
lp/mm acrysof
12
acrysof vivity/luxsmart
12
vivity/luxsmart crystal
12
crystal aperture
12
depth focus
8
intraocular lenses
8
edof iols
8

Similar Publications

A Comparative Analysis of the Effects of Misaligning Different Trifocal Intraocular Lenses.

J Clin Med

December 2024

The David J Apple Center for Vision Research, Department of Ophthalmology, Heidelberg University Eye Clinic, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany.

This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea.

View Article and Find Full Text PDF

Wide FOV metalens for near-infrared capsule endoscopy: advancing compact medical imaging.

Nanophotonics

November 2024

National Key Laboratory on Near-Surface Detection, School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China.

This study presents the design, fabrication, and characterization of a wide field-of-view (FOV) metalens optimized for capsule endoscopy. The metalens achieved a 165° FOV with a high modulation transfer function (MTF) of 300 lines per millimeter (lp/mm) across the entire FOV, operating in the near-infrared (NIR) narrow-bandpass imaging at 940 nm. The performance of the metalens-based system is evaluated using two bandwidths, 12 nm and 32 nm, showing MTF values of 0.

View Article and Find Full Text PDF
Article Synopsis
  • Wide-angle metalenses in the longwave infrared can outperform traditional lenses because they are lighter, more affordable, and compatible with CMOS technology.
  • This study introduces two metasurfaces to correct common optical issues like distortion and spherical aberration, resulting in a hybrid metalens designed for thermal imaging in the 8-12 μm waveband.
  • The new metalens shows impressive performance metrics, such as an FOV of 24°, an F-number of 1.2, and has potential applications in areas like camera technology, autonomous vehicles, healthcare, and environmental monitoring.
View Article and Find Full Text PDF
Article Synopsis
  • The study explores the development of a linac-mounted photon-counting detector (PCD) for image-guided radiotherapy (IGRT) that could significantly enhance imaging by improving soft-tissue contrast and spatial resolution compared to traditional flat panel detectors (FPDs).
  • Researchers characterized the image quality of the PCD, focusing on parameters like 2D spatial resolution, noise, and contrast, by comparing it with an FPD using various imaging techniques and calibration methods.
  • Results showed that the PCD provides a linear energy calibration and better raw contrast in images, suggesting its potential for improved tumor delineation in radiation therapy applications.
View Article and Find Full Text PDF

The geostationary orbit imaging spectrometer offers distinct advantages for diverse applications in remote sensing. To enhance swath and data richness, there is a growing trend toward spectrometers with broader fields of view (FOV) and extended slit lengths. Optical systems equipped with these features face significant challenges in aberration correction.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!