The detection limit of lateral flow immunoassay (LFIA) is largely determined by the properties of the label used. We compared four nanoparticle labels differing in their chemical composition and colour: (1) gold nanoparticles (Au NPs), red; (2) Au-core/Pt-shell nanoparticles (Au@Pt NPs), black; (3) latex nanoparticles (LPs), green; and (4) magnetic nanoparticles (MPs), brown. The comparison was carried out using one target analyte-, the causal bacterial agent of fire blight. All nanoparticles were conjugated with antibodies through methods that provide maximum functional coverage like physical adsorption (Au NPs, Au@Pt NPs) and covalent bonding (LPs, MPs). All conjugates demonstrated the same ability to bind with through enzyme-linked immunosorbent assay where optical properties of the nanoparticles do not determine the registered signal. However, half-maximal binding was achieved at different numbers of nanoparticles because they differ in size. All conjugates based on four nanoparticle labels were used for lateral flow assays. As a result, Au@Pt NPs provided the minimal detection limit that corresponded to 10 CFU/mL. Au NPs and LPs detected 10 CFU/mL, and MPs detected 10 CFU/mL. The results highlight that simply choosing a coloured label can significantly affect the detection limit of LFIA.