Residual host cell proteins (HCPs) in the drug product can affect product quality, stability, and/or safety. In particular, highly active hydrolytic enzymes at sub-ppm levels can negatively impact the shelf life of drug products but are challenging to identify by liquid chromatography-mass spectrometry/mass spectrometry (LC-MS/MS) due to their high dynamic range between HCPs and biotherapeutic proteins. We employed new strategies to address the challenge: (1) native digest at a high protein concentration; (2) sodium deoxycholate added during the reduction step to minimize the inadvertent omission of HCPs observed with native digestion; and (3) solid phase extraction with 50% MeCN elution prior to LC-MS/MS analysis to ensure effective mAb removal. A 50 cm long nanoflow charged surface hybrid column was also packed to allow for higher sample load for increased sensitivity. Our workflow has increased the sensitivity for HCP identification by 10- to 100-fold over previous reports and showed the robustness as low as 0.1 ppm for identifying HCPs (34.5 to 66.2 kDa MW). The method capability was further confirmed by consistently identifying >85% of 48 UPS-1 proteins (0.10 to 1.34 ppm, 6.3 to 82.9 kDa MW) in a monoclonal antibody (mAb) and the largest number (746) of mouse proteins from NIST mAb reported to date by a single analysis. Our work has filled a significant gap in HCP analysis for detecting and demonstrating HCP clearance, in particular, extremely low-level hydrolases in drug process development.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1021/acs.analchem.1c03095 | DOI Listing |