Background: Mobile applications which are designed to assess the range of motion (ROM) are widely used.

Objective: The aim of this cross-sectional observational study was to determine the inter-observer and intra-observer reliability of a smartphone application "PT Goniometer" (PTG) and determine the correlation between PTG and universal goniometer (UG) regarding active ROMs of the hip in healthy participants.

Methods: Thirty-four healthy young participants were included in the study. Two physiotherapists performed active hip flexion, abduction, internal rotation and external rotation ROM measurements of dominant legs of the participants by using PTG and UG. Intraclass correlation coefficients (ICC) were calculated to determine the intra-observer and inter-observer reliability. Level of correlations between PTG and UG were used to establish concurrent validity of PTG.

Results: The PTG demonstrated excellent inter-observer and intra-observer reliability (ICC > 0.90) for all measured hip movements. The minimum detectable change (MDC95) was ranged from 3.29° to 5.1° for the intra-observer reliability, and from 2.55° to 3.21° for the inter-observer reliability. Additionally, the concurrent validity was found excellent (r = 0.91-0.93).

Conclusion: The results of the present study suggest that PTG is a valid and reliable mobile technology for measuring hip ROMs.

Download full-text PDF

Source
http://dx.doi.org/10.3233/WOR-213626DOI Listing

Publication Analysis

Top Keywords

intra-observer reliability
12
reliability smartphone
8
measuring hip
8
inter-observer intra-observer
8
inter-observer reliability
8
concurrent validity
8
ptg
6
hip
5
reliability
5
validity reliability
4

Similar Publications

Accurate quantitative assessments are crucial to understanding development of diseases and their effective treatments. Various validated perimetry and volumetry measurement methods for patients with lymphedema exist and each has its own advantages and limitations and choosing the right instrument is essential. PeriKit® (PK) is a new measurement device that requires validation.

View Article and Find Full Text PDF

AI-Assisted Compressed Sensing Enables Faster Brain MRI for the Elderly: Image Quality and Diagnostic Equivalence with Conventional Imaging.

Int J Gen Med

January 2025

School of Biomedical Engineering & State Key Laboratory of Advanced Medical Materials and Devices, ShanghaiTech University, Shanghai, People's Republic of China.

Purpose: Conventional brain MRI protocols are time-consuming, which can lead to patient discomfort and inefficiency in clinical settings. This study aims to assess the feasibility of using artificial intelligence-assisted compressed sensing (ACS) to reduce brain MRI scan time while maintaining image quality and diagnostic accuracy compared to a conventional imaging protocol.

Patients And Methods: Seventy patients from the department of neurology underwent brain MRI scans using both conventional and ACS protocols, including axial and sagittal T2-weighted fast spin-echo sequences and T2-fluid attenuated inversion recovery (FLAIR) sequence.

View Article and Find Full Text PDF

Assessment of the interobserver and the intraobserver reproducibility for the detection of renal cortical defects in adults and children using [99mTc]Tc-MAG3.

Q J Nucl Med Mol Imaging

January 2025

Section of Nuclear Medicine and Diagnostic Imaging, International Atomic Energy Agency, Vienna, Austria.

Background: One can assess cortical defects on the early images of [99mTc]Tc-MAG3 renography. We aimed to assess interobserver and intraobserver reproducibility for detecting renal cortical defects using [99mTc]Tc-MAG3 for adults and children; identify causes for poor inter- and intraobserver reproducibility and to assess the effect of the kidney to background ratio (KTBR) on reproducibility.

Methods: One hundred adult and 200 pediatric renograms were included.

View Article and Find Full Text PDF

Purpose: To investigate image quality and agreement of derived cardiac function parameters in a novel joint image reconstruction and segmentation approach based on disentangled representation learning, enabling real-time cardiac cine imaging during free-breathing.

Methods: A multi-tasking neural network architecture, incorporating disentangled representation learning, was trained using simulated examinations based on data from a public repository along with MR scans specifically acquired for model development. An exploratory feasibility study evaluated the method on undersampled real-time acquisitions using an in-house developed spiral bSSFP pulse sequence in eight healthy participants and five patients with intermittent atrial fibrillation.

View Article and Find Full Text PDF

: The aim of this study is to assess the inter- and intra-observer reproducibility of the identification of direct and indirect ultrasonographic features of adenomyosis as defined by the revised Morphological Uterus Sonographic Assessment (MUSA) consensus (2022). : A cohort of 74 women, aged 18 to 45, were recruited from the recurrent miscarriage and general gynaecology clinic at a university-based fertility centre. All the participants underwent 2D and 3D transvaginal Ultrasound scan (TVS) examination in the late follicular and early luteal phase.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!