Objectives: The body composition monitor (BCM) is a bioimpedance spectroscopy device, specifically developed for patients on hemodialysis (HD) to improve ultrafiltration (UF) programming, based on an objective assessment of the degree of overhydration (OH) at the start of HD. However, its acceptance in clinical practice remains limited because of concerns about the accuracy at the individual level. The aim of this study is to examine the performance of the BCM and to identify means of improvement.

Methods: Precision of the OH estimate was assessed by 6 consecutive measurements in 24 patients on HD. Accuracy was examined in 45 patients, by comparing the change in OH (ΔOH) during HD with UF volume. Accuracy was considered acceptable if the volume error in individual patients was ≤0.5 L.

Results: The OH estimate had an analytical precision of 1.0 ± 0.4%. The correlation between UF volume and ΔOH was moderate (Slope = 0.66, R = 0.44, P < .001) and indicated underestimation of UF volume, in particular for high UF volumes. Accuracy at individual level was highly variable. A volume error >0.5 L occurred in 44% of patients. Accuracy improved over the course of HD, with a decrease in total error range from 2.3 L in the first hour to 1.1 L in the final hour of HD.

Conclusions: The accuracy of BCM volume change estimates is highly variable and below requirements of daily practice. Improvement may be achieved by a switch to an end-of-HD measurement.

Download full-text PDF

Source
http://dx.doi.org/10.1053/j.jrn.2021.11.004DOI Listing

Publication Analysis

Top Keywords

bioimpedance spectroscopy
8
patients hemodialysis
8
patients accuracy
8
accuracy
6
patients
6
accuracy bioimpedance
4
spectroscopy detection
4
detection hydration
4
hydration changes
4
changes patients
4

Similar Publications

Purpose: This study compared %BFUS to %BF4C in young adult athletes.

Methods: University club sport athletes (86 women, 138 men) from a variety of teams participated. ADP, DXA, and bioimpedance spectroscopy were used to measure body volume, bone mineral content, and total body water, respectively, for the 4C model.

View Article and Find Full Text PDF

: Hemodialysis patients face a high mortality risk, requiring effective clinical assessments. In these patients, muscle wasting due to protein-energy wasting (PEW) leads to increased frailty, which is strongly associated with worse outcomes, including higher mortality. As muscle mass declines, so does functional capacity, making regular assessment of both muscle mass and function critical for prognostic evaluation.

View Article and Find Full Text PDF

A comprehensive review of computational diagnostic techniques for lymphedema.

Prog Biomed Eng (Bristol)

January 2025

Amrita Vishwa Vidyapeetham, Center for Wireless Networks & Applications (WNA), Amrita Vishwa Vidyapeetham Amritapuri, Kollam, India, Kollam, 690525, INDIA.

Lymphedema is localized swelling due to lymphatic system dysfunction, often affecting arms and legs due to fluid accumulation. It occurs in 20% to 94% of patients within 2 to 5 years after breast cancer treatment, with around 20% of women developing breast cancer-related lymphedema (BCRL). This condition involves the accumulation of protein-rich fluid in interstitial spaces, leading to symptoms like swelling, pain, and reduced mobility that significantly impact quality of life.

View Article and Find Full Text PDF

The placenta plays a critical role in nutrient and oxygen exchange during pregnancy, yet the effects of medicinal drugs on this selective barrier remain poorly understood. To overcome this, this study presents a cost-effective bioimpedance spectroscopy (BIS) system to assess tight junction integrity and monolayer formation in BeWo b30 cells, a widely used model of the multinucleated maternal-fetal exchange surface of the placental barrier. Cells were cultured on collagen-coated porous membranes and treated with forskolin to induce controlled syncytialization.

View Article and Find Full Text PDF

Prototype analysis of a low-power, small-scale wearable medical device.

J Electr Bioimpedance

January 2024

Electrical Engineering Department, State University of Santa Catarina, Santa Catarina, Brazil.

Wearable and portable devices are gaining significant popularity across consumer electronics as well as in medical and industrial fields. To ensure that these devices are both comfortable and appealing to users, they need to have low battery consumption and be compact in both size and weight. The EGluco project is focused on developing a wearable device for non-invasive blood glucose monitoring.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!