This study aims to quantify the stresses of sonographers using two different ultrasound devices, one of conventional and one of ergonomic design. A total of 20 obstetricians and gynecologists participated in this study, and two types of tasks (scanning and positioning) were evaluated while using the two different devices. To quantify workload, four dependent variables (muscle activity, estimated grip force, subjective comfort rating, and task time) were measured. The muscular activity required while using the conventional device was 14.4% MVC (Maximum voluntary contraction) for the scanning task, which was significantly higher than that of the ergonomic device. The subjective comfort rating for the conventional design was lower than that of the ergonomic design. For the positioning task, the ergonomic device (33.2% MVC) resulted in significantly higher muscle activity in the extensor digitorum (ED) and flexor digitorum superficialis (FDS) than the conventional design (22.2% MVC), whereas the deltoid muscle showed significantly lower activity than in users of conventional design (4.5% MVC). Ergonomically-designed ultrasound devices improve ease of moving and the probe's supporters, reduce physical load and increase ease of use for sonographers. Our results may be used as guidelines for usability testing of ultrasound devices.