Background: This study aimed to establish a robust, simple method to detect 25-hydroxyvitamin D (25(OH)D ), 25-hydroxyvitamin D (25(OH)D ), 1,25-dihydroxyvitamin D (1,25(OH) D ), 1,25-dihydroxyvitamin D (1,25(OH) D ), 24,25-dihydroxyvitamin D (24,25(OH) D ), and 24,25-dihydroxyvitamin D (24,25(OH) D ) simultaneously with efficient separation of 3-epi 25(OH)D , 3-epi 24,25(OH) D , 23R,25(OH) D , and 4β,25-dihydroxyvitamin D (4β,25(OH) D ) by liquid chromatography-tandem mass spectrometry (LC-MS/MS).
Method: This method was validated according to procedures established by Clinical and Laboratory Standards Institute (CLSI) and then applied in healthy population to determine the distribution of the vitamin D metabolites by LC-MS/MS.
Results: The total-run CV% of 25(OH)D , 25(OH)D , 24,25(OH) D , 24,25(OH) D , 1,25(OH) D , and 1,25(OH) D were 6.30%-8.40%, 5.00%-8.40%, 5.90%-9.00%, 5.60%-9.00%, 5.60%-8.00%, and 7.00%-9.70%, respectively. The linearity correlation coefficients r of these six vitamin D metabolites were >0.99. The matrix effects of 25(OH)D , 25(OH)D , 24,25(OH) D , 24,25(OH) D , 1,25(OH) D , and 1,25(OH) D were 90.6%-103.3%, 97.3%-106.3%, 90.7%-106.3%, 100.7%-114.5%, 97.9%-104.6%, and 97.0%-111.0%. The trueness values of 25(OH)D , 25(OH)D , and 24,25(OH) D were 93.8%-103.0%, 101.0%, and 96.3%-100%, respectively.
Conclusion: This study successfully established an efficient, accurate, robust method for simultaneous measurement of serum 25(OH)D, 1,25(OH) D, and 24,25(OH) D by LC-MS/MS with efficient separation of 3-epi analogs, 23R,25(OH) D , and 4β,25(OH) D .
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jms.4792 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!