Neurostimulation with multiple scalp electrodes has shown enhanced effects in recent studies. However, visualizations of stimulation-induced internal current distributions in brain is only possible through simulated current distributions obtained from computer model of human head. While magnetic resonance current density imaging (MRCDI) has a potential for direct in-vivo measurement of currents induced in brain with multi-electrode stimulation, existing MRCDI methods are only developed for two-electrode neurostimulation. A major bottleneck is the lack of a current switching device which is typically used to convert the DC current of neurostimulation devices into user-defined waveforms of positive and negative polarity with delays between them. In this work, we present a design of a four-electrode current switching device to enable simultaneous switching of current flowing through multiple scalp electrodes.

Download full-text PDF

Source
http://dx.doi.org/10.1109/EMBC46164.2021.9630962DOI Listing

Publication Analysis

Top Keywords

current switching
12
switching device
12
device enable
8
magnetic resonance
8
current
8
resonance current
8
current density
8
density imaging
8
multiple scalp
8
scalp electrodes
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!