Introduction: Silicone radial head prostheses (SRHP) have a reputation of acceptable initial function but failure in the medium to long term as a result of the disintegration of the silicone material. Damaged SRHP can result in a silicone-induced synovitis and destruction of the joint. Early removal of damaged SRHP may prevent joint destruction. Literature is scarce; there are no studies describing early symptoms associated with clinical failure of SRHP and very few studies on appropriate treatment of destructed SRHP. The aim of this study was to describe the clinical course of failing SRHP and to provide practical guidance for treatment.
Methods: Fifteen consecutive cases of failed SRHP, operated between 2005 and 2020, were retrospectively reviewed, using patient logs, radiographic results, and a questionnaire. Relevant data concerning pre- and postoperative complaints, function, and surgical details were gathered. Using the questionnaire, retrospective and current patient reported outcomes including Single Assessment Numeric Evaluation (SANE) scores, pain, instability, and range of motion were analyzed.
Results: The average time between implantation of the SRHP until outpatient presentation was 16 years. The average recalled SANE score before start of complaints was 8.6 out of 10 points. In 4 patients, the onset of complaints was preceded by a trauma. Five patients had symptoms for more than a year, and 9 patients for several months. Progressive pain in the elbow and/or wrist and mechanical symptoms were early and prominent symptoms of failure. Clinical signs of synovitis were frequently present. The majority of patients had a slow but progressive onset of complaints over several months before presentation. Twelve of 15 cases of failed SRHP were surgically treated. Fragmentation of the implant, osteoarthritis, and synovitis were found in all surgically treated cases. Both surgical removal and synovectomy alone, and revision to a new radial head prosthesis resulted in good medium-term outcomes with a mean SANE score of 7.3 points at 8.4 years of follow-up. Postoperative extension lags and mild instability were frequently mentioned as a reason for reimplantation of a metallic radial head implant and removal of the radial head alone, respectively.
Conclusion: Both patients and physicians should be aware of early symptoms of a failing SRHP. Adequate recognition, radiographic evaluation, and early intervention may favor outcomes of failing SRHP. Surgical removal is easy to perform. In selected cases, revision of the radial head prosthesis may be considered. Both of these treatment strategies resulted in satisfying outcomes.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.jse.2021.10.043 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!